4.6

Settings

The ICC-2000 is supplied with a number of standard settings that can be adjusted by the user:
The constants used in the CO
and/or correlation with other reference standards.
The desired values for pressure, temperature and CO
Values at which the instrument is to be calibrated can be indicated.
4.7

Reading and operating

Display
The ICC-2000 has a TFT display and three keys ("↑", "↓" and "START"). The display is activated
as soon as one of the keys is used. After the last operation, the screen will remain switched on
for a variable programmable length of time.
Attention!
The display indicates:
Software version number immediately when switched on ("<
Values measured for pressure ("P") and temperature ("T"),
Value calculated for the CO
Whether a value is outside the measuring range ("HHH" or "LLL"),
When a battery needs charging ("BATTERY"),
Number under which the measurement is stored
Whether the value measured of pressure or temperature is stable (stops flashing)
Operating levels
The ICC-2000 has 4 levels can be set to the various operating levels. The four operating levels
are not secured.
Measuring screen
Configuration
Calibration
Logbook
See the "Software-Summary" in Appendix VII.
ICC
formula can be refined by the user for specific beverages
2
Wait until the values stop increasing before reading. If the display
is flashing, the value shown is unstable and still changing.
content ("CO2"),
2
For reading the values measured and calculated.
To set values, date/time/display, formula parameters and
language.
To calibrate pressure and check the temperature and pressure.
To view stored results and calibrations.
HAFFMANS ICC
content.
2
V 1.00
>"),
Page 15 of 49