9.

MAINTENANCE AND CLEANING

Attention!
9.1

ICC-2000 maintenance (semi-annual)

Procedure of replacing the seals (needle valve and temperature sensor):
1. Unscrew the needle valve (pos. 13) from the instrument,
2. Remove the o-ring (pos. 14) and the nylon ring (pos. 15) from the needle valve,
3. Put a new nylon ring on the needle valve, then a new o-ring.
4. Screw the needle valve back into the instrument.
5. Loosen the socket head screws on the bottom of the measuring head next to the
needle,
6. Replace all o-rings (pos.nr. 25, 65),
1. Assemble the bottom of the measuringhead.
9.2

Charging the battery

Attention!
The ICC-2000 has a built-in chargeable NiMH battery. A fully charged battery can be used for
about 5 hours. If the battery is dead, "BATTERY" will flash on the display and the battery should
then be charged using the charging station supplied. Charging the battery takes about 10 hours.
The batteries will be charged whether the ICC-2000 is switched on or off. The ICC-2000 can
also be charged using a connected charger. The battery is protected against overcharging, so
that the ICC-2000 can remain safely connected to the mains.
When charging:
The charger's low-voltage plug should be inserted in the feed connector at the back of
the ICC-2000,
The charger's feed plug should be in a mains socket.
ICC
For optimal maintenance of ICC-2000 the Inpack 2000 Service set
is recommended.
Before connecting, check that the charger's connection is suitable
for the mains voltage.
HAFFMANS ICC
Page 31 of 49