4.
Wait until normal monitoring screen appears, then check that the picture on the screen is
displayed with correct resolution.
If the resolution is not correct, replace the remote controller cable.
?
5.
Enter the service menu:
Menu
(on the remote controller keypad)
4-34) - Service (password 26-23-8)
Take down the information regarding remote controller software.
?
Select the menu KEYBOARD:
Service - Keyboard
6.
Check the remote controller keys.
Press the keys one by one. Check that each key generates a sound from the loudspeaker.
?
7.
Check the ComWheel.
Turn the ComWheel clockwise and counterclockwise and check that each step generates a
sound and the corresponding values at the bottom of the menu increase.
Select DUMMY PRESS. Press the ComWheel and check that the press generates a sound and
the corresponding value in the menu increases.
?
8.
Perform electrical safety check and leakage current test.
?
9.
Check that the remote controller functions normally after the performed electrical safety
check.
?
10. Clean the remote controller and the cable.
?
Fill in all necessary documents.