Installing oxygen
inlet connector and
AC power cord
The Respironics V60/V60 Plus Ventilator destined for Japan, China, and the
USA are typically pre-configured. V60/V60 Plus Ventilators shipped to other
countries may require installation of the power cord and oxygen inlet
connector.
1. Install the oxygen inlet connector as follows (Figure A-4):
a. Gently fit connector into the hole provided with flat sides to the
left and right.
b. Install the oxygen inlet connector retaining plate. Tighten the two
screws with a 2.5-mm hex wrench.
Oxygen
inlet
connector
Retaining plate
Figure A-4: Installing the oxygen inlet connector
First-time installation
A-7