Table 7-1: HFT Alarm and other messages: summary and troubleshooting
Message
Description
Cannot Reach Target
Displays when HFT (high
Flow
flow therapy) is active. In-
dicates that flow target is
not achieved.
Patient alarms are
Displays when HFT (high
disabled during HFT
flow therapy) is active. Pa-
tient alarms are not avail-
able in this therapy.
Patient Circuit Oc-
Displays when HFT (high
cluded
flow therapy) is active.
Gas flow to the patient is
obstructed.
Check Vent: Proxi-
Displays when HFT (high
mal Pressure Sensor
flow therapy) is active.
Proximal pressure is out of
Range Error
range.
Corrective Action
Check the patient. Check
that the high flow nasal
cannula size is appropri-
ate for the flow setting.
Check that an occlusive
interface is NOT in use (a
cannula fully sealed with-
in the nares, an NIV mask
or direct connection to an
ETT/trach). Check for an
occlusion, kink or liquid in
the patient circuit.
Manually reset to confirm
and clear the audible
alarm.
Check the patient. Check
that an occlusive interface
is NOT in use (a cannula
fully sealed within the na-
res, an NIV mask or direct
connect to an ETT/trach).
Check for an occlusion,
kink or liquid in the pa-
tient circuit. If problem
persists, provide alterna-
tive ventilation.
When using a high flow
nasal cannula with FEP
Connect, the proximal
pressure line should be
disconnected from the
ventilator port during HFT.
If this message is seen
when switching the pa-
tient from ventilation to
HFT, check to make sure
the proximal pressure line
is disconnected.
If the message still oc-
curs, provide alternative
ventilation and have the
ventilator serviced.
High flow therapy
Priority
Manually
Autores
type (ID)
resettable
ettable
Low
No
Yes
(66)
Low/
Yes
No
Infor-
mation
(68)
High
No
Yes
(67)
High
Yes
No
(12)
Silence
able
Yes
Yes
Yes
No
7-7