Technical specifications
Physical
characteristics
Environmental
specifications
11-6
Table 11-7: Physical characteristics
Parameter
Dimensions
Mass (weight) of the
V60/V60 Plus ventilator in its
most usual configuration
Installed weight
(V60/V60 Plus ventilator on
stand, including accessories as
listed)
Maximum load
(Mass/weight of the V60/V60
Plus ventilator stand, including
its safe working load)
Table 11-8: Environmental specifications
Parameter
Temperature
Relative humidity
Barometric pressure
Specification
(33.7 cm)
13.3 in.
(39.4 cm)
15.5 in.
12 kg (26 lb) with battery
V60/V60 Plus Ventilator with backup battery, ventila-
tor stand, O
tank holder, two E-cylinders (full), two
2
gauge clusters, oxygen analyzer with mount, humidifi-
er with chamber, water bag and pole, nebulizer, circuit
arm with mount, patient circuit with mask, O
fold with hoses, remote alarm cable, and power cord.
Weight: 64 kg (142 lb)
70.5 kg (155 lb)
Specification
Operating: 5 to 40 ºC (41 to 104 ºF)
Storage/transport: -20 to 50 ºC (-4 to 122 ºF)
Operating: 15 to 95% (noncondensing)
Storage/transport: 10 to 95% relative
(noncondensing)
Operating: 600 mmHg to 765 mmHg (80 kPa to
102 kPa): approximately -61 m to 1951 m (-200 ft to
6400 ft) relative to sea level
Storage/transport: 450 mmHg to 765 mmHg (60 kPa
to 102 kPa): approximately -61 m to 4267 m
(-200 ft to 14000 ft) relative to sea level
(42.9 cm)
16.5 in.
mani-