Alarms, messages, and troubleshooting
Table 9-3: Alarm and other messages: summary and troubleshooting (continued)
Message
Description
Proximal Pressure
Proximal pressure low for
Line Disconnect
a few seconds. Proximal
pressure line is discon-
nected. Air flow to the pa-
tient continues.
Running on Internal
System is powered by the
Battery
internal battery. Autore-
sets when ventilator is
connected to AC power.
Trigger:+x,
Auto-Trak+ is active and
E-Cycle: +x
using the displayed set-
Use Menu to change
tings. This messages is
displayed for 5 min after
start-up.
Using Default Set-
Displayed after power on
tings
if setting values are cor-
rupted or not set, or if de-
fault values were restored
by the user.
Vent Inoperative x
See Table 9-5 on page 9-16
description of failure
Table 9-4: Check Vent alarm messages: summary and troubleshooting
Message
Description
Check Vent: 1.8 V
Technical failure
Supply Failed
Check Vent: 3.3 V
Technical failure
Supply Failed
Check Vent: 5 V Sup-
Technical failure
ply Failed
Check Vent: 12 V
Technical failure
Supply Failed
Check Vent: 24 V
Technical failure
Supply Failed
9-12
Corrective Action
Check the patient. Recon-
nect proximal pressure
line. Confirm ventilator
and alarm settings are ap-
propriate. If problem per-
sists, provide alternative
ventilation. Have the ven-
tilator serviced.
Connect ventilator to AC
power.
Confirm that Auto-Trak+
settings are appropriate.
Check the patient. Check
and adjust settings as re-
quired.
Corrective Action
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Check the patient. Provide
alternative ventilation. Have
the ventilator serviced.
Priority
Manually
Autores
type (ID)
resettable
ettable
High
Yes
Yes
(41)
Low
Yes
Yes
(60)
Infor-
Yes
Yes
mation
(65)
Infor-
Yes
Yes
mation
(61)
Priority
Manually
Autores
type (ID)
resettable
ettable
High
Yes
No
(21)
High
Yes
No
(22)
High
Yes
No
(23)
High
Yes
No
(24)
High
Yes
No
(25)
Silence
able
Yes
Yes
N/A
N/A
Silence
able
No
No
No
No
No