Removing/Installing The Front Shroud/Battery; Retention Bracket; Batteries - Invacare TDX SP User Manual

Hide thumbs Also See for TDX SP:
Table of Contents

Advertisement

Invacare® TDX® SP

8.8.3 Removing/Installing the Front Shroud/Battery

Retention Bracket

Removing
The front shroud/battery retention bracket C is removed from and
installed onto the wheelchair frame D as a single unit.
Front shroud/battery retention bracket C mounts on hooks E
located on the wheelchair frame D.
Have the following tools available: 1/4 inch Allen Wrench,
7/16 inch Box Wrench.
1.
Remove the mounting screws F securing the front
shroud/battery retention bracket to the wheelchair frame D.
2.
Lift up to unhook the front shroud/battery retention bracket
from the base frame and remove.
Installing
Wheelchairs with TRRO or TRBKTS Only - Battery
retention brackets MUST be installed at all times.
When performing STEP 1, ensure that the tabs G located on
the sides of the front shroud/battery retention bracket C
are positioned on the hooks E located on the front of the
wheelchair frame D.
1.
Reinstall the front shroud/battery retention bracket onto front
of wheelchair frame.
Secure the front shroud/battery retention bracket C to the
2.
wheelchair frame D with two mounting screws F. Securely
tighten.
62
3.
Gently, pull on the front shroud to ensure the mounting screws
are fully engaged into the battery box.
4.
Wheelchairs with Footboards Only: Using the mounting screw
H and washer I, secure the footboard support J to the
wheelchair frame.

8.9 Batteries

WARNING!
Risk of Death or Serious Injury
Failure to observe these warnings can cause an electrical
short resulting in death, serious injury, or damage to
the electrical system.
– The POSITIVE (+) RED battery cable MUST connect
to the POSITIVE (+) battery terminal(s)/post(s).
– The NEGATIVE (-) BLACK battery cable
MUST connect to the NEGATIVE (-) battery
terminal(s)/post(s).
– NEVER allow any of your tools and/or battery cable(s)
to contact BOTH battery post(s) at the same time.
An electrical short may occur and serious injury or
damage may occur.
– Install protective caps on positive and negative battery
terminals.
– Replace cable(s) immediately if cable(s) insulation
becomes damaged.
– DO NOT remove fuse or mounting hardware from
POSITIVE (+) red battery cable mounting screw.
1143190-N-03

Advertisement

Table of Contents
loading

Table of Contents