Invacare TDX SP User Manual page 63

Hide thumbs Also See for TDX SP:
Table of Contents

Advertisement

WARNING!
Risk of Injury or Damage
Improper installation of the battery can result in injury
or damage.
– Batteries can weigh up to 52 lbs (23.6 kg). ALWAYS
use a battery lifting strap when lifting the battery. It
is the most reliable method of carrying a battery and
preventing serious injury.
WARNING!
Risk of Injury or Damage
Improper lifting technique may cause injury or damage.
– Use proper lifting techniques, assistance and gear such
as straps when available when lifting heavy loads.
WARNING!
Risk of Injury
Exposure to battery acid may result in injury.
– The use of rubber gloves is recommended when
working with batteries.
– DO NOT allow the liquid in the battery to come in
contact with skin, clothes or other possessions. It is
a form of acid and harmful or damaging burns may
result. Should the liquid touch your skin, wash the
area IMMEDIATELY and thoroughly with cool water.
In serious cases or if eye contact is made, seek medical
attention IMMEDIATELY.
– DO NOT install/reinstall a battery with a cracked or
otherwise damaged case.
1143190-N-03
Invacare strongly recommends battery installation and
replacement be performed by a qualified technician.
Read the installation instructions noted on the battery and
in the manual.
Ensure there is no battery acid in the bottom or around the
battery box or on the sides of the battery(ies). Neutralize
battery acid with baking soda if found.
Clean battery tray and batteries prior to installation.
Use battery lifting strap when available.
Keep batteries in an upright position and avoid tipping.
Use only deep cycle gel cell battery.
Use box wrench when tightening clamps. Do Not wiggle the
battery terminals/posts when tightening.
Setup and Maintenance
63

Advertisement

Table of Contents
loading

Table of Contents