IBA ibaPADU-S-IT-2x16 Manual

Central unit for the iba modular system
Table of Contents

Advertisement

Quick Links

ibaPADU-S-IT-2x16
Central unit for the iba modular system
Manual
Issue 1.9
1
Issue 1.9

Advertisement

Table of Contents
loading

Summary of Contents for IBA ibaPADU-S-IT-2x16

  • Page 1 Central unit for the iba modular system Manual Issue 1.9 Issue 1.9...
  • Page 2 However, the information in this publication is updated regularly. Re- quired corrections are contained in the following regulations or can be downloaded on the Inter- net. The current version is available for download on our web site www.iba-ag.com. Copyright notice Windows ®...
  • Page 3: Table Of Contents

    Application examples ................... 20 8.1.1 Measuring systems with ibaPDA ..............20 8.1.2 Measuring system with ibaPDA and ibaLogic-V5 ......... 20 8.1.3 Standalone systems with ibaLogic-V5............21 8.1.4 ibaPADU-S-IT-2x16 in hierarchical automation systems ......21 FO cascading ....................21 Issue 1.9...
  • Page 4 Network settings directly on the device ............27 Time settings ....................30 9.3.1 System time ....................30 9.3.2 Time synchronization of the local ibaPADU-S-IT-2x16 system ..... 31 Make settings on the Website ..............33 9.4.1 Accessing the Website ................33 9.4.2 Website structure ..................
  • Page 5 Manual 12.4 Dimensions ....................73 12.5 Connection diagram ..................80 12.5.1 Pin assignment voltage supply X14 ............. 80 12.5.2 Pin assignment digital inputs X5 ..............80 12.6 Example for FO budget calculation .............. 81 Accessories and related products............... 83 Appendix ......................
  • Page 6: About This Manual

    About this manual In this manual, you learn a lot about the design of the ibaPADU-S-IT-2x16 device and how to use and operate it. You can find a general description of the iba-modular system and further information about the design of the modules and how to use and operate them in separate manuals.
  • Page 7: Target Group

    Manual Target group This manual addresses in particular the qualified professionals who are familiar with han- dling electrical and electronic modules as well as communication and measurement tech- nology. A person is regarded to as professional if he/she is capable of assessing safety and recognizing possible consequences and risks on the basis of his/her specialist training, knowledge and experience and knowledge of the standard regulations.
  • Page 8: Used Symbols

    Manual ibaPADU-S-IT-2x16 Used symbols If safety instructions or other notes are used in this manual, they mean: The non-observance of this safety information may result in an imminent risk of death or severe injury:  By an electric shock!  Due to the improper handling of software products which are coupled...
  • Page 9: Introduction

    CPU, but also up to 4 input/output modules. The ibaPADU-S-IT- 2x16 CPU offers 8 digital inputs. ibaPADU-S-IT-2x16 is a fast CPU for fast applications in the fields of measurement tech- nology, signal processing and controls. When equipped with the right input/output mod- ules and combined with the ibaPDA or ibaLogic-V5 software products, ibaPADU-S-IT- 2x16 can be used for several applications.
  • Page 10: Safety Instructions

    Never supply the device with a voltage other than 24 V DC ±10%! A higher voltage may destroy the device or be dangerous to life! Modules and CPU must NOT be attached or detached to/from the rack under voltage! Switch off ibaPADU-S-IT-2x16 or disconnect power supply before attaching/detaching the modules. Important note...
  • Page 11: System Requirements

    Software  ibaPDA beginning with version 6.37.3  ibaLogic beginning with version 5.0.3 Note An ibaLogic-V5 embedded version is running on an ibaPADU-S-IT-2x16 device by default. The license is included in delivery. ibaPADU-S-IT-2x16  Firmware beginning with version v02.14.017  Hardware beginning with version A0...
  • Page 12: Mounting, Connecting, Dismounting

    Manual ibaPADU-S-IT-2x16 Mounting, Connecting, Dismounting Works on the device must NOT be done when it is under voltage! Mounting 1. Mount the backplane on an appropriate construction. 2. Connect the ground terminal. 3. Attach the device to the left slot.
  • Page 13: Device Description

    Manual Device description Views Operating status indicators L1 … L4 On/off switch S11 Connector for power supply 24 V X14 Status LED digital inputs L10 … L17 Connection for digital inputs X5 Fixing screws Indicators L5 … L8 Network interface X22...
  • Page 14: Indicator Elements

    Only if the green system LED L1 flashes evenly and none of the LEDs L5 ... L8 is on, the device is ready for operation or can be switched off properly. Important note When the LED L4 indicates a failure, please contact the iba Support. 7.2.2 LEDs L5 … L8 The LEDs L5 …...
  • Page 15: Status Of Digital Inputs L10

    For further information, see chapters  9.1.4 “Set back to default settings“  9.2.4 “Network settings directly on the device“  11.1.7 “Deleting an ibaLogic-V5 program in ibaPADU-S-IT-2x16” Issue 1.9...
  • Page 16: System Function Push Button S10

     X11 (RX): FO input  X10 (TX): FO output Process data between ibaPADU-S-IT-2x16 and the connected iba systems are trans- ferred via FO cables. Using the 32Mbit Flex transfer protocol, also configuration data can be transferred via FO cable.
  • Page 17: Debounce Filters

    Manual Figure 4: Connection diagram for digital inputs X5 7.5.2 Debounce filters For the digital inputs, there are four debounce filters for each. These can be chosen and configured for each signal independently. You have got the following filters at your dis- posal: ...
  • Page 18 Manual ibaPADU-S-IT-2x16 „Stretch rising edge“ With the first rising edge, the input signal (red) switches to logical 1 and keeps this value for the defined debounce time. Thereafter, the channel is transparent again and waits for the next rising edge.
  • Page 19: Voltage Supply

    Manual „Delay both edges“ Beginning with the first edge, the output signal (purple) blocks the input and keeps the logical value of the edge for the duration of the defined debounce time. Thereafter, the channel is transparent again, directly assumes the logical level of the input signal and waits for the next edge –...
  • Page 20: System Integration

    Manual ibaPADU-S-IT-2x16 System integration Application examples The figures below show examples of ibaPADU-S-IT-2x16 combined with ibaPDA and ibaLogic-V5. 8.1.1 Measuring systems with ibaPDA Figure 9: Measuring system with ibaPDA  Pure measuring applications (capturing, recording, analyzing)  Recording the signals with ibaPDA 8.1.2...
  • Page 21: Standalone Systems With Ibalogic-V5

     Master control program on the ibaLogic PC  Fast local control  Fast deterministic communication via ibaNet to the central computer FO cascading With 32Mbit Flex, you can group up to 15 modular systems to a ring topology. Figure 13: Ring topology with ibaPADU-S-IT-2x16 Issue 1.9...
  • Page 22: Supported Ibanet Transfer Protocols

    Manual ibaPADU-S-IT-2x16 The devices are addressed via the S1 rotary switch Device number in the Position of rotary switch S1 cascade Not allowed 1st device 2nd device 14th device 15 th device A cascade can be formed with all devices that support the ibaNet 32Mbit Flex protocol.
  • Page 23: Initial Start-Up

    You can find the necessary data on the name plate. The default host name is made up of the device type and the serial number. Example: Device type: ibaPADU-S-IT-2x16 Serial number: 000004  Host name: S-IT2x16-000004 Figure 14: name plate...
  • Page 24: Set Back To Default Settings

    Manual ibaPADU-S-IT-2x16 9.1.4 Set back to default settings If you have changed the network settings or the passwords, and want to set them back to default values, please proceed as follows: 1. Switch off the device. 2. Position the rotary switches S1 on „6“ and S2 on „9“.
  • Page 25 IP address via „Control panel – Network connections“, e.g. 192.168.1.2 Note The IP address must not be 192.168.1.1, as this is the address of ibaPADU-S-IT-2x16. Use this IP address for accessing the Websites. 3. Start a Web browser on the PC and insert the IP address of the device in the URL: 192.168.1.1...
  • Page 26: Establishing Communication To The Device Via Fo Cable

    Manual ibaPADU-S-IT-2x16 9.2.2 Establishing communication to the device via FO cable We recommend this connection if the device has a 32Mbit Flex FO connection to ibaPDA. 1. Start ibaPDA and open the I/O Manager.  See chapter 11.2.1 “Configuration in the I/O Manager“...
  • Page 27: Network Settings On The Web Interface

    Manual 9.2.3 Network settings on the Web interface Important note If you do not get access to the Website under Windows 7 despite a connection is es- tablished, please change the security settings. See note in chapter 9.4.1 "Accessing the Website".
  • Page 28 Manual ibaPADU-S-IT-2x16 Procedure for setting a new fixed address: 1. Switch off the device by pressing S11. 2. Switch the rotary switches S1 to „F“ and S2 to „E". 3. Push the S10 push button and keep it pressed (firmly up to the stop point) and then switch on the device again.
  • Page 29 Manual In the following example, the network address 192.168.1.1 is set for the network interface X22: 1. L5 flashing Position S1 on „C “and S2 on „0“ and then press S10. 1. Octet number „192“ is set. 2. L5 lighted, L6 flashing: Set S1 on „A “and S2 on „8“...
  • Page 30: Time Settings

    You can make the time settings only on the website. Important note ibaPADU-S-IT-2x16 cannot store the time without external buffer voltage, if no supply voltage is connected to X14. If the time should be buffered, refer to chapter 7.6.2 “Buffer voltage X30”.
  • Page 31: Time Synchronization Of The Local Ibapadu-S-It-2X16 System

     NTP protocol  DCF77 NTP protocol You can set the ibaPADU-S-IT-2x16 internal system time via the NTP protocol, provided that an Ethernet connection (X22) is established to the Internet or a local NTP server. Figure 16: Time synchronization with NTP protocol 1.
  • Page 32 3. Choose “Obtain system time from external signal“ 4. Then, make the following settings:  “Input Channel“ Channel number of the ibaPADU-S-IT-2x16 digital input channel to which the time signal is connected.  “clock polarity“ Choose the value of the digital input signal that corresponds to the active state.
  • Page 33: Make Settings On The Website

    Other documentation Moreover, an additionally connected ibaPDA measuring system can be synchronized via any digital input (e.g. ibaPADU-S-IT-2x16 or a digital input module), please refer to ibaPDA manual. Make settings on the Website The Website offers functions for diagnostics and settings of the device like passwords, network settings, time settings, firmware update.
  • Page 34 Parameters: Network security: LAN Manager authentification level "Send LM & NTLM responses" Value: You will find the registry entry on the „iba Software & Manuals“ DVD. Other documentation There is a module specific website for each I/O module. These are described in the man- uals for the single modules.
  • Page 35: Website Structure

    Manual 9.4.2 Website structure The website contains the following tabs: Functions padu admin info Displaying general information about the device firmware Displaying current firmware eventlog Displaying system events passwords Displaying passwords and possibility to change them network Displaying network settings and changing them...
  • Page 36 Manual ibaPADU-S-IT-2x16 9.4.2.2 “firmware“ tab On the “firmware“ tab, you can find information about the current firmware version. Figure 21: “firmware“ tab Note The “Overall release version“ refers to the whole ibaPADU-S system, including the mod- ules. When you get and install a new version, the CPU and all modules are updated.
  • Page 37 Manual Figure 23: “passwords“ tab 1. If you want to change a password, please enter it into the fields. For safety purposes, please enter the old password once and the new password twice. 2. Then, please click on <change>.
  • Page 38 Manual ibaPADU-S-IT-2x16 settings, please click on the <submit> button. If you have made a mistake when typing the address, please click on the <reset entries> button. The settings are reset to the last valid version. The fields below displays the MAC address of the device.
  • Page 39 Manual When using Internet Explorer 7 or earlier, the following security settings in IE are neces- sary, so that the backup function can be used. Select under Extras – Internet Optionens - Security and click on <Custom level>. The “Security Settings”...
  • Page 40 <disable>: deactivating autostart  Autostart image Displaying if an ibaLogic-V5 program is stored on ibaPADU-S-IT-2x16, which can be started automatically triggered by autostart of the PMAC. 9.4.2.11 “notes“ tab On the “notes“ tab, you can enter notes, e.g. for notes on wiring or on recording of changes.
  • Page 41: Updates

    Now the update is completed and the device is restarted automatically. When installing a firmware update, the entire iba modular system is updated, i.e. the CPU and the I/O modules attached to the system.
  • Page 42: Update Via Ibapda

    After having mounted the modules and applied the voltage to the central unit, ibaPADU- S-IT-2x16 detects the modules and checks the firmware version. ibaPADU-S-IT-2x16 has a so called “overall release version“, i.e. the firmware version of the entire system. This version contains the current firmware version of the central unit as well as the firmware versions of the modules.
  • Page 43 Web interface. In this case, a new update file has to be installed for the “overall release version“. If you want to get the current update file, please contact the iba support. Issue 1.9...
  • Page 44: Iba Applications

    11.1 ibaLogic-V5 You can use ibaPADU-S-IT-2x16 – in cooperation with ibaLogic-V5 – for the following applications: individual signal processing, control applications or stand-alone applica- tions. ibaLogic-V5 programs can be loaded in the device as runtime versions and run there autonomously.
  • Page 45 If there are other projects in this workspace, you need to “activate“ the new project. Configure ibaPADU-S-IT-2x16 as platform via the Tools – Platform Configuration menu. Figure 33: Start platform configuration For creating or configuring a platform, please click in the dialog under the project name on <New item>...
  • Page 46 An alternative way for creating a platform is to open the Current Platform dropdown menu and select “Add platform”. Figure 36: Choose “Add platform” in the menu line If ibaPADU-S-IT-2x16 is configured as platform, yet, you can choose the platform via the dropdown box in the menu line. Issue 1.9...
  • Page 47: Automatic Update Of The Ibalogic-V5 Version

    11.1.2 Automatic update of the ibaLogic-V5 version The application verifies when configuring an ibaPADU-S-IT-2x16 platform or starting an ibaLogic-V5 program, whether the ibaLogic-V5 runtime version of ibaPADU-S-IT-2x16 is compatible to the local ibaLogic-V5 version. If not, you are asked automatically to update the ibaPADU-S-IT-2x16.
  • Page 48: Configuring Signals

    On the “Hardware Configuration“ tab, you can make the settings for ibaLogic-V5 and all available resources. If ibaPADU-S-IT-2x16 is configured as platform, only ibaPADU-S-IT-2x16 is offered as interrupt source. Under “time base“, you set the smallest possible cycle time. The “Enable Watchdog“...
  • Page 49 Manual Under “Module Settings“, ibaPADU-S-IT-2x16 is always activated. If “Buffered Access“ is enabled, the imported values are buffered and supplied to the program as arrays. The smallest possible interval of ibaLogic-V5 is 1 ms. If you want to measure signal values with sampling rates smaller than 1 ms, you need to record the signal values “buffered“.
  • Page 50 Manual ibaPADU-S-IT-2x16 Signal Meaning Invalid_In_ X[2..5] Status signals for service purposes Invalid_Out_X[2..5] Invalid_In_FO Invalid_Out_FO Invalid_In_App Invalid_Out_App Err_Sample_Clk Err_App_Crash Outputs DebType_Ch[00…07] Debounce filter for digital signals DebTime_Ch[00...07] Debounce time for the single digital signal SampleTime Sample time, especially for buffered access and...
  • Page 51 If combined with ibaPDA, “External“ must be chosen. In this case, the FiberOptics_IO module of ibaPADU-S-IT-2x16 is displayed as virtual ibaPADU-S module in ibaPDA. This is also the default setting. If “External“ is chosen as setting for the input, “External“...
  • Page 52 For the “External“ Link protocol, the S1 switch needs to be positioned on 1-F. For all other settings S1 = 0. Important note When using the 32Mbit link protocol, iba recommends to configure the selected 32Mbit timebase with the SampleTime output in addition to ensure a cycle-synchronous trans- mission via FO.
  • Page 53 Figure 41: Modular measuring system with intelligent pre-processing In the example above, “External” has to be chosen as FO protocol for input and output. The communication between ibaPADU-S-IT-2x16 and ibaPDA is done in the 32Mbit Flex mode. Figure 42: ibaLogic-V5 as module in ibaPDA...
  • Page 54 Stand-alone system with ibaLogic-V5: When ibaPADU-S-IT-2x16 is used as stand-alone system with ibaLogic-V5 runtime, the measured values may be saved in iba measuring files by means of a DAT_FILE_WRITE (DFW) function block, see chapter 11.1.4 “DAT_FILE_WRITE function block”. Note After having done all signal settings, and before doing the “Assign signal“...
  • Page 55: Dat_File_Write Function Block

    11.1.4 DAT_FILE_WRITE function block Using a DAT_FILE_WRITE (DFW) function block data can be stored in the RAM or Flash memory of ibaPADU-S-IT-2x16. For this purpose the DFW function block has to be con- figured accordingly. Other documentation For a detailed description of the DFW function block and the usage please refer to the ibaLogic-V5 manual.
  • Page 56 Manual ibaPADU-S-IT-2x16 Figure 46: DAT_FILE_WRITE (DWF) function block The measuring files can be stored either on the RAM (Random Access Memory, RAM drive) or the flash memory (solid state drive) of the device. Both memory devices differ in size, access and how they are addressed.
  • Page 57: Configuring The Debounce Filter

    Manual The flash memory is preferably intended for the acquistion of not time sensitive and slow signals, usually for continuous recordings. The flash memory provides with 110 GB sufficient memory size. Access to the flash memory: \\<IP address or hostname of ibaPADU-S-IT>\FlashDisk\...
  • Page 58: Saving The Ibalogic-V5 Program On Runtime

    Manual ibaPADU-S-IT-2x16 Stretch falling edge Stretch both edges Delay both edges  You find the explanations of the settings in chapter 7.5.2 “Debounce filter“. Now, connect the OTC and the signal on the margin of the programming surface. Create a new OTC, assign a meaningful name, e.g. debounce time and choose “Input”...
  • Page 59: Deleting An Ibalogic-V5 Program In Ibapadu-S-It-2X16

    PMAC in online mode. If Autostart is enabled on the „ibaLogic“ tab of the ibaPADU-S-IT-2x16 Website, the ibaLogic-V5 program is started automatically after you have switched on the device.
  • Page 60: Ibapda

    Start ibaPDA, open the I/O Manager and proceed as follows: 1. Look in the I/O Manager for the link of the FOB-D card, ibaPADU-S-IT-2x16 is con- nected to. Click with the right mouse button on the link. The submenu is opened.
  • Page 61 5. Click on the “General” tab on “Read configuration from device“. The connected modules are detected automatically and shown in the signal tree. 6. Configure ibaPADU-S-IT-2x16 and the modules (e.g. assign a name, debouncing, etc.) (see chapter 11.2.2 „PADU-S – General tab“ ff.).
  • Page 62: Padu-S - General Tab

    If TRUE is selected, the module name is placed in front of the signal name as pre- fix. Connection  IP address The IP address or the host name of the ibaPADU-S-IT-2x16 device (only for infor- mation).  Auto enable/disable When this option is enabled and ibaPDA cannot connect to this device during the start of the acquisition then it will disable this module and start the acquisition with- out it.
  • Page 63: Padu-S - Diagnostics Tab

     Write firmware Using this button you can install a firmware update. Select the update file „ padu- sit2x16_v[xx.yy.zzz].iba“ in the browser and start the update with <OK>. Important note The update may take several minutes and must not be interrupted. After an update the device will be automatically rebooted.
  • Page 64: Padu-S - Analog Tab

    Manual ibaPADU-S-IT-2x16 Note The function “Reset to factory defaults” does not apply to the reset of network settings or passwords as described in chapter 9.1.4. Network settings and passwords remain stored.  After the reset the following message appears and the device will be automatically re-initialized with the deleted I/O settings.
  • Page 65: Ibapadu-S-It-2X16 - General Tab

    Manual Figure 56: PADU-S module – “Digital“ tab 11.2.6 ibaPADU-S-IT-2x16 – General tab Figure 57: ibaPADU-S-IT-2x16 module – “General“ tab Basic settings  Module Type, Locked, Enabled, Name, Timebase see chapter 11.2.2 “PADU-S – General tab “  Module No.
  • Page 66: Ibapadu-S-It-2X16 - Digital Tab

     Configuration set by This item is only visible when ibaPADU-S-IT-2x16 is used as central unit. When an embedded application has been started on ibaPADU-S-IT-2x16 (e. g. ibaLogic), then ibaPDA cannot modify the configuration of the modules and signals. In this case the configuration is set by the embedded application.
  • Page 67: Diagnostics Signals

    Manual falling edge, stretch both edges, delay both edges.  See chapter. 7.5.2 „Debounce filters“.  Debounce time (µs) Here, you can define the debounce time in µs  Enable Enabling/disabling the signal Note Configure the plugged analog and digital modules. For a detailed description, please read the manuals referring to the modules.
  • Page 68 Manual ibaPADU-S-IT-2x16 11.2.8.1 Diagnostics – General tab Figure 60: Diagnostics module – “General” tab Basic settings  Module Type, Locked, Enabled, Name, Module No., Use name as prefix see chapter 11.2.2 “PADU-S – General tab “  Timebase The timebase is related to the general acquisition timebase of the ibaPDA system.
  • Page 69 Manual The diagnostics signals can be activated individually. The meaning of the signals: Signal Meaning Hardware state Indicates whether the IO module is ready for operation. Hardware available Indicates whether the IO module has been detected and initialized properly during startup.
  • Page 70: Technical Data

    Manual ibaPADU-S-IT-2x16 Technical Data 12.1 Main data Short description Name ibaPADU-S-IT-2x16 Description Central unit for the iba modular system Order number 10.124021 Processor unit Processor 1.6 GHz Atom processor, dual core CPU Operating system Windows Embedded Compact 7 Main memory...
  • Page 71: Interfaces

    47 CFR § 2.1077 Compliance Information Unique Identifier: 10.124021 ibaPADU-S-IT-2x16 10.124000 ibaPADU-S-B4S Responsible Party - U.S. Contact Information iba America, LLC 370 Winkler Drive, Suite C Alpharetta, Georgia 30004 (770) 886-2318-102 www.iba-america.com FCC Compliance Statement This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference re- ceived, including interference that may cause undesired operation.
  • Page 72: Digital Inputs

    Manual ibaPADU-S-IT-2x16 Further interfaces Ethernet 10/100 Mbit/s 2x host, 1x device for service purposes 12.3 Digital inputs Digital inputs Number Design Electrically isolated, protected against reverse polarity, single ended Input signal 24 V DC Max. signal voltage ±60 V permanent Signal range log.
  • Page 73: Dimensions

    Manual 12.4 Dimensions ibaPADU-S-IT-2x16 (Dimensions in mm) Figure 62: Dimensions ibaPADU-S-IT-2x16 (Dimensions in mm) Figure 63: Dimensions ibaPADU-S-IT-2x16 with cables Issue 1.9...
  • Page 74 Manual ibaPADU-S-IT-2x16 Distance between 2 ibaPADU-S-IT-2x16 systems (Dimensions in mm) Figure 64: Min. distance between 2 ibaPADU-S-IT-2x16 systems Issue 1.9...
  • Page 75 Manual ibaPADU-S-IT-2x16 and backplane (Dimensions in mm) Figure 65: Dimensions ibaPADU-S-IT-2x16 with backplane Issue 1.9...
  • Page 76 Manual ibaPADU-S-IT-2x16 Mounting plate 19“ (Dimensions in mm) Figure 66: Dimensions of mounting plate Issue 1.9...
  • Page 77 Manual Backplane ibaPADU-S-B4S (Dimensions in mm) Figure 67: Dimensions ibaPADU-S-B4S Backplane ibaPADU-S-B4S with mounting angles (Dimensions in mm) Figure 68: Dimensions ibaPADU-S-B4S with mounting angles Issue 1.9...
  • Page 78 Manual ibaPADU-S-IT-2x16 Backplane ibaPADU-S-B1S for one central unit and one module (Dimensions in mm) Figure 69: Dimensions ibaPADU-S-B1S (Dimensions in mm) Figure 70: Dimensions ibaPADU-S-B1S with modules Issue 1.9...
  • Page 79 Manual Subrack ibaPADU-S-B for one central unit (Dimensions in mm) Figure 71: Dimensions ibaPADU-S-B Issue 1.9...
  • Page 80: Connection Diagram

    Manual ibaPADU-S-IT-2x16 12.5 Connection diagram 12.5.1 Pin assignment voltage supply X14 Connection + 24 V 12.5.2 Pin assignment digital inputs X5 Connection Digital input 00 + Digital input 00 - Digital input 01 + Digital input 01 - Digital input 02 +...
  • Page 81: Example For Fo Budget Calculation

    Manual 12.6 Example for FO budget calculation As an example, an FO connection from an ibaFOB-io-Dexp card (FO transmitter) to an ibaBM-PN device (FO receiver) is used. Figure 72: Example: simple connection for FO budget calculation The example refers to a point-to-point connection with an FO cable of type 62.5/125 µm.
  • Page 82 Manual ibaPADU-S-IT-2x16 Equation for calculating the FO budget (A Budget ���� = |(���� − ���� ������������������������ ���� ���������������� ���� �������� ���� �������������������� = sensitivity of FO receiving interface Receiver = output power of FO transmitting interface Sender Equation for calculating the fiber optic cable length (l ����...
  • Page 83: Accessories And Related Products

    Manual Accessories and related products Backplane ibaPADU-S-B4S Order number 10.124000 Backplane panel (for carrying components) for 1 central unit and up to 4 I/O modules w x h x d: 9.02 in x 8.62 in x 0.83 in (229 mm x 219 mm x 21 mm)
  • Page 84 Manual ibaPADU-S-IT-2x16 Mounting panel 19“ for PADU-S modular Order number 10.124005 Mounting panel (483 mm/19“) for up to two ibaPADU-S-B4S backplanes Mounting 1 ibaPADU-S-B4S centered or 2 ibaPADU-S-B4S left and right Mounting equipment included Module Carrier for ibaPADU-S modular system Order number 10.124007...
  • Page 85 30.772048 For up to 2048 signals Online signal processing ibaLogic-V5 embedded ibaLogic-V5 license already existing on ibaPADU-S-IT-2x16 for 64 inputs/outputs and Dat- FileWrite license ibaAnalyzer-V7 33.010000 Offline- and online analysis software with free li- cense if used to analyze *.dat files generated by li- censed iba software.
  • Page 86: Appendix

    Manual ibaPADU-S-IT-2x16 Appendix 14.1 Acronyms DHCP Dynamic Host Configuration Protocol Daylight Saving Time Fiber Optics Greenwich Mean Time Network Time Protocol PMAC Programmable Measurement and Automation Controller Real Clock Time TCP/IP Transmission Control Protocol / Internet Protocol Universal Time Coordinated...
  • Page 87: Index

    Manual 14.2 Index Network settings directly on the device 15, 16, 17, 28 Backplane on the Web interface Website Notes Communication to the device via FO cable 27 entering notes Communication to the device via the network25 NTP protocol...
  • Page 88: Support And Contact

    Mailing address iba AG Postbox 1828 D-90708 Fuerth Germany Delivery address iba AG Gebhardtstrasse 10 D-90762 Fuerth Germany Regional and Worldwide For contact data of your regional iba office or representative please refer to our web site www.iba-ag.com. Issue 1.9...