12 Care and maintenance

12.1 Maintenance instructions

• Put a dust cover over the instrument while the breaks are in
work.
• Keep accessories in a dust-free place when not in use.
• Remove dust with a pneumatic rubber pump and a soft brush.
• Clean the objectives and eyepieces with special optics cleaning
cloths and pure alcohol.
• Protect the surgical microscope from damp, vapors, acids,
alkalis, and corrosive substances.
Do not keep chemicals near the instrument.
• Protect the surgical microscope from improper handling.
Install other device sockets or unscrew optical systems and
mechanical parts only when explicitly instructed to do so in this
user manual.
• Protect the surgical microscope from oil and grease.
Never oil or grease the guide surfaces or mechanical parts.
• Remove coarse debris with a moistened disposable cloth.
• To disinfect the surgical microscope, use compounds from the
surface disinfectant group based on the following active
ingredients:
• aldehydes,
• alcohols,
• quaternary ammonium compounds.
Due to potential damage to the materials, never use
products based on
• halogen-splitting compounds,
• strong organic acids,
• oxygen-splitting compounds.
Follow the disinfectant manufacturer's instructions.
X
It is recommended to conclude a service contract with Leica
Service.

12.2 Cleaning the touch panel

Before cleaning the touch panel, switch off your PROVEO 8 and
X
disconnect it from the power supply.
Use a soft, lint-free cloth to clean the touch panel.
X
Do not apply cleaning agent directly to the touch panel; rather,
X
apply it to the cleaning cloth.
Use a commercially available glass/eyeglass cleaner or plastic
X
cleaner to clean the touch panel.
Do not apply pressure to the touch panel while cleaning it.
X
It is recommended to conclude a service contract with Leica
Service.
CAUTION
Damage of the touch panel!
X
X

12.3 Maintenance

The PROVEO 8 surgical microscope generally requires no
maintenance. To ensure that it always operates safely and reliably,
we recommend that you take the precaution of contacting the
responsible service organization.
You can arrange for periodic inspections or, if appropriate, conclude
a maintenance contract with them.
• It is recommended to conclude a service contract with
• Use only original spare parts for servicing.
• After 18 months you will be reminded that the
Press the "CONFIRM" button.
X
The dialog window is closed.
PROVEO 8 / Ref. 10 733 910 / Version 02
Operate the touch panel using your fingers only.
Never use hard, sharp or pointed objects made out of
wood, metal or plastic.
Never clean the touch panel using cleaners that contain
abrasive substances. These substances can scratch the
surface and cause it to be become dull.
Leica Service.
inspection is due when you switch on the microscope.
Care and maintenance
61