SYMPTOM
CPR MONITOR
CPR Monitor lights
are not working
properly
To help understand how
the CPR Monitor works,
please note that the CPR
Monitor lights are
powered by a small
current generated by the
CPR Monitor
Please follow the
troubleshooting tips at
right in the order they
are listed for best
results
CPR Monitor lights are
all on
CPR Monitor red
light is blinking
MANIKIN TORSO AND FACE-SHIELD/LUNG-BAG
Cannot hear a beep
when doing
compressions
COPYRIGHT© 2021 PRESTAN PRODUCTS, LLC
TROUBLESHOOTING GUIDE
POSSIBLE ISSUES OR COMMENTS
•
Chest compressions are not deep
enough to signal CPR Monitor
•
Batteries are not installed, need
replacing
or
are
correctly in the CPR Monitor
•
CPR Monitor may be clogged with
plastic dust, interrupting the signal
to the Monitor
•
CPR Monitor needs to be replaced
• Batteries are running low
• This is normal after replacing batteries,
manikin is paused or is connecting to
CPR Feedback app ‐ may blink up to 10
minutes to indicate ready status
•
Hand position and/or compression
angle may be incorrect
•
Compressions may not be deep
enough to activate the CPR Monitor
•
CPR Monitor may be broken
•
Batteries
may
be
misassembled
•
Compress the chest until the beep is
audible from the CPR Monitor at 2"‐
2.4"
•
not inserted
Install or replace batteries into CPR
Monitor (back of manikin, left
shoulder area) ‐ requires 2‐AA (1.5V)
batteries
•
Remove CPR Monitor and blow out
dust – then replace CPR Monitor and
reconnect
•
Exchange CPR Monitor with working
manikins (if available) to determine if
CPR Monitor needs to be replaced
•
•
Commence or resume compressions ‐
blinking will stop and feedback rate
signals will begin
•
Confirm correct hand placement and
compress straight down from directly
above the manikin
•
Compress between 2 and 2.4 inches
depth
•
Replace batteries
dead
or
21
SOLUTIONS
Replace batteries
12716-RevB