SYMPTOM
Unit does not turn on
Both pads attached to
manikin but device will not
advance to analysis
Device proceeds to
analyzing before both pads
have been placed on the
manikin
Device delivers shock
before shock button is
pressed
Device prompts "Replace
Batteries"
Device suddenly shuts
down or voice prompts are
distorted
Pad pulls out from pad
connector
Pad will not stick to the
manikin
Grooves/indentations are
appearing on the surface
of the pads
COPYRIGHT© 2020 PRESTAN PRODUCTS, LLC
TROUBLESHOOTING GUIDE
POSSIBLE ISSUES OR COMMENTS
• Unit must be manually turned on
• Batteries are not property installed
• Batteries are dead or are missing
• Operational solution
• Pad is inserted upside down in the cable
connector
• Pad or Adhesive worn out
• Pads not properly positioned on tray
• Look for pad riding up pad tray wall
• Device is In Automatic mode
• Device will run 10 or more hours with low
batteries
• Batteries are very low, almost dead
• Check for black connector pin missing
• Adhesive is worn out
• The cables are indenting into the surface
of the pads from improper storage
placement
•
Push green power button
•
Re-orient batteries properly
•
Replace batteries
•
Insure pads are pressed on firmly,
specifically so that the sensor in the
center of the adhesive is touching a
surface
•
Push out black connector pin, flip
pad around and re-secure with
black connector pin
•
Replace pad (lasts 100+ uses)
•
Re-position pad on tray to lay
completely flat so it is not touching
side walls
•
Change back to Semi-Automatic
Mode
•
Replace batteries within 10 hours
•
Install new batteries
•
Install new black connector pin
•
Replace pads (lasts 100+ uses)
•
Store cables and pads in their
respective pouches in the interior
lid of the unit's carry bag
9
SOLUTIONS