SETUP

Setup mode allows viewing and modifying the following parameters:
• Salinity Factor
• Altitude Factor
• Log Interval
• Current Time (hour & minute)
• Current Date (month, day & year)
• Beep Status
• Instrument ID
• Temperature Unit
To enter the Setup mode press SETUP while the instrument is in mea‑
surement mode. Press SETUP again to exit SETUP mode.
Select a parameter with the ARROW keys.
Press CAL if you want to change a parameter value. The selected
parameter will blink.
Press RANGE to toggle between displayed paramet+ers.
Press the ARROW keys to increase or decrease the displayed value.
Press CFM to save the modified value or CAL to escape.
Press the ARROW keys to select the next/previous parameter.
SALINITY FACTOR
Press CAL when the salinity factor is displayed. The salinity factor ("0"
to "40" g/l) and the "CFM" tag will start blinking.
Press ARROW keys to change the salinity factor value.
Press CFM to save the modified value or press CAL to escape without
saving.
The salinity affects the DO concentration, decreasing its value. The next
table shows the maximum oxygen solubility at various temperatures
and salinity levels.
14