Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipping. Thank you for choosing a Hanna Instruments Product. If there is any damage, notify your dealer. Please read this instruction manual carefully before using the instru- Each meter is supplied complete with: ment.
OPERATIONAL GUIDE PRELIMINARY OPERATIONS TAKING RESISTIVITY MEASUREMENTS • Plug the supplied 12 Vdc adapter into the power supply socket • Immerse the HI 3316D probe in the solution, on the rear panel. while making sure that the hole on the probe body is completely submerged.
Page 5
CONDUCTIVITY VS. TEMPERATURE CHART CALIBRATION °C °F HI 7030 HI 7031 HI 7033 HI 7034 HI 7035 HI 7039 (μS/cm) (μS/cm) (μS/cm) (μS/cm) (μS/cm) (μS/cm) CONDUCTIVITY CALIBRATION 7150 7 7 6 48300 65400 2760 • It is recommended to calibrate the instrument for the conductivity 8220 8 9 6 53500...
DETERMINATION OF THE TEMPERATURE CONDUCTIVITY PROBE MAINTENANCE COEFFICIENT OF A SOLUTION • Immerse the probe into a sample of the solution and adjust the Rinse the probe with tap water after conductivity TC knob to 0% (i.e. no compensation). every series of measurements. •...
ACCESSORIES CONDUCTIVITY CALIBRATION SOLUTIONS PROBES HI 7030M 12880 μS/cm, 230 mL bottle HI 76303 Platinum 4-ring conductivity probe with built- in temperature sensor, DIN connector and 1 m HI 7030L 12880 μS/cm, 500 mL bottle (3.3') cable HI 8030L 12880 μS/cm, 500 mL FDA bottle HI 3316D Resistivity probe with built-in temperature sen- HI 7031M...
If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization number from the Customer Service department and then send it with shipping costs prepaid.