DESCRIPTION or high surface tension liquids. It is a quality product from OPERATING RANGES AND MATERIALS Gilson, which is fully compliant ISO 8655, CE labeled. FITTING THE CAPILLARY AND PISTON The six models available allow for precise and accurate SETTING THE VOLUME pipetting of viscous, dense and volatile liquids.
2 - PARTS CHECK LIST 4 - OPERATING RANGES AND MATERIAL Available Models Just take a moment to verify that the following items are present: Model Volume Range MICROMAN E, M10E 1 µL to 10 µL User's Guide, M25E 3 µL to 25 µL M50E...
Make sure the capillary piston (CP) button to the first stop until you feel and hear a slight MICROMAN E model matches the MICROMAN E click 3. model you are about to use: CP10 M10E 4) Do not release the push-button and pipette your solution.
You may be able to identify and correct the problem (top position), by referring to the following table. If you cannot solve - Wipe any liquid from the out- the problem, please contact your Gilson representative. side of the capillary, taking care Leaks: not to touch the orifice.
Chemical Decontamination Cleaning Clean the pipette prior to decontamination. Full details Clean the pipette with soap solution prior to decon- of recommended decontamination procedures for Gilson tamination. pipettes are available from your supplier. Whichever other External decontaminant you use, check with the supplier that the...
The figures provided in the (FD10002) ± 0.27 0.08 ± 0.7 0.3 “ Gilson Maximum Permissible Errors ” table were obtained ± 0.30 0.10 ± 0.7 0.3 using Gilson CPs. These figures are only guaranteed when M50E Min.
Gilson or its designates. Gilson’s sole liability www.gilson.com under this warranty shall be limited to, at Gilson’s sole option, repair or replacement of any defective components of pipettes or refund of the Hereby certifies under its sole responsibility that the products purchase price paid for such pipettes.