Guide Lines for Protection against Radiation

The X-ray system may cause injury to the patients if used improperly. The instructions contained in this
manual must be read and followed when operating PaX-Primo. The world standard regulations pertaining
to radiation safety must be observed.
When exposing X-ray, User must be behind the protective wall, or take other protective actions. When a
breakdowns or troubles appear, User keeps at least 2m (7feet) away from the X-ray system to release the
exposure switch while observing patient and capture-progress.
User must provide the protective clothes to the patient. Before capturing, pregnant women must always
consult with doctors.
Do not use this system in any place where there is an explosive danger.

X-ray Room Space for System Installation

Recommended Minimum Space

PaX-Primo
- Above space is considered for the movement of both system operator and patient.
- The system is normally installed beside a wall, and operator uses the system on left.

Width of Door

- The width of door is more than 800mm for system movement into X-ray room.
Ground
- The condition of ground should be flat for system balance.
- Ground should support a min. 500KG of weight.
Power Supply
- For system stability please guarantee 3KW power supply.
Protection against radiation
- For protection against radiation follow the government or local standards.
Safety Zone
- Check the safety zone.
2000(L) * 2000(W) * 2500(H) mm