TX Level

To adjust the transmitter output level of the LT51 press the [f1] Tx Level key.
Press the [f1] softkey until 600 is displayed for the ZRTX and ZTX impedances.
Terminate the LT51 TXa-b terminals with an external 600  load. Using a short,
low capacitance cable connect the TXa-b terminals to the high impedance input
of a DMM configured to measure ACV and display the measurement in dBm
referenced to 600 .
Use the knob on the LT51 to adjust the level displayed on the DMM to 0.00 dBm
and then press the
NOTE: The knob must be turned (even if the default TxLevel is correct ) to zero
the calibration constant, before pressing the
value. If the knob is turned without saving the value, power cycle the
LT51 before continuing.
When the 600  Tx Level has been adjusted, press the [f1] softkey to adjust the
Tx Level for 75  and 50 . For each impedance follow the procedure outlined
above for the 600  adjustment, but with the appropriate termination. For these
two impedances, the level ranges of the LT51 are not adjusted, since the values
obtained during the 600  adjustment are sufficient.
LT51 User Manual
key.
Adjustment procedures
key to save the new
78