Making Measurements

DC / AC Voltage

Measure the voltage between the RTXa-b terminals:
press the
key and then press [f2] to select Voltage, DC/AC.
The DC and AC voltage between the RTXa-b terminals is continuously
monitored and displayed on the LCD.
The AC voltage displayed is the true RMS value of the signal detected at the
selected RTX terminals. It does not include any DC component of the signal.
A high impedance (1 M) is always used to make the Voltage measurements
and the Measurement Interface is always set to TX-RX.
DC voltages less than 0.5V are displayed as 0.00V.
LT51 User Manual