Summary of Contents for Endress+Hauser Proline 500 PROFINET
Page 1
Products Solutions Services KA01348D/06/EN/01.17 71376021 Brief Operating Instructions Proline 500 PROFINET Transmitter with electromagnetic sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Transmitter Brief Operating Instructions Contain information about the transmitter. Sensor Brief Operating Instructions →...
Page 2
Proline 500 PROFINET Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
Page 3
These Brief Operating Instructions are the Transmitter Brief Operating Instructions. The "Sensor Brief Operating Instructions" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer...
Proline 500 PROFINET Document information Document information Symbols used 1.1.1 Safety symbols Symbol Meaning DANGER! DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING! This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in WARNING serious or fatal injury.
Page 6
Document information Proline 500 PROFINET Symbol Meaning Protective Earth (PE) A terminal which must be connected to ground prior to establishing any other connections. The ground terminals are situated inside and outside the device: • Inner ground terminal: Connects the protectiv earth to the mains supply.
Proline 500 PROFINET Basic safety instructions Basic safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task. ‣ Are authorized by the plant owner/operator.
Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
Proline 500 PROFINET Basic safety instructions Device-specific IT security The device offers a range of specific functions to support protective measures on the operator' s side. These functions can be configured by the user and guarantee greater in-operation safety if used correctly.
Product description Proline 500 PROFINET Product description The measuring system consists of a Proline 500 transmitter and an electromagnetic Proline Promag sensor. The transmitter and sensor are mounted in physically separate locations. They are interconnected via two connecting cable. Transmitter with integrated ISEM (intelligent sensor electronics module)
Proline 500 PROFINET Installation Installation For detailed information about mounting the sensor, see the Sensor Brief Operating Instructions → 3 Mounting the transmitter housing CAUTION Ambient temperature too high! Danger of electronics overheating and housing deformation. ‣ Do not exceed the permitted maximum ambient temperature .
Page 12
Installation Proline 500 PROFINET 4.1.2 Post mounting WARNING Order code for "Transmitter housing", option L "Cast, stainless": cast transmitters are very heavy. They are unstable if they are not mounted on a secure, fixed post. ‣ Only mount the transmitter on a secure, fixed post on a stable surface.
Proline 500 PROFINET Installation Turning the transmitter housing To provide easier access to the connection compartment or display module, the transmitter housing can be turned. – – 3 mm 4 mm A0029993 Depending on the device version: Loosen the securing clamp of the connection compartment cover.
Installation Proline 500 PROFINET Turning the display module The display module can be turned to optimize display readability and operability. – – – 3 mm A0030035 Depending on the device version: Loosen the securing clamp of the connection compartment cover.
Proline 500 PROFINET Installation Cover locking NOTICE Order code "Transmitter housing", option L "Cast, stainless": The covers of the transmitter housing are provided with a borehole to lock the cover. The cover can be locked using screws and a chain or cable provided by the customer.
Installation Proline 500 PROFINET Transmitter post-installation check The post-installation check must always be performed after the following tasks: • Mounting the transmitter housing: – Post mounting – Wall mounting • Turning the transmitter housing • Turning the display module Is the device undamaged (visual inspection)? ...
Proline 500 PROFINET Electrical connection Electrical connection NOTICE The measuring device does not have an internal circuit breaker. ‣ For this reason, assign the measuring device a switch or power-circuit breaker so that the power supply line can be easily disconnected from the mains.
Page 18
Electrical connection Proline 500 PROFINET Signal cable PROFINET Standard IEC 61156-6 specifies CAT 5 as the minimum category for a cable used for PROFINET. CAT 5e and CAT 6 are recommended. For more information on planning and installing PROFINET networks, see: "PROFINET Cabling and Interconnection Technology", Guideline for PROFINET...
Page 19
Proline 500 PROFINET Electrical connection Cable length (max.) Depends on the medium conductivity, max. 200 m (656 ft) Cable lengths (available for 5 m (15 ft), 10 m (32 ft), 20 m (65 ft) or variable length up to max. 200 m (656 ft)
Page 21
Proline 500 PROFINET Electrical connection 5.1.7 Preparing the measuring device Carry out the steps in the following order: Mount the sensor and transmitter. Connection housing, sensor: Connect connecting cable. Transmitter: Connect connecting cable. Transmitter: Connect signal cable and cable for supply voltage.
Page 22
Electrical connection Proline 500 PROFINET Preparing the connecting cable: Promag H Transmitter Electrode cable Coil current cable 70 (2.76) 80 (3.15) 50 (1.97) 17 (0.67) 50 (1.97) 8 (0.3) 10 (0.4) 8 (0.3) A0029544 A0029543 Engineering unit mm (in) A = Terminate the cable B = Fit ferrules on cables with fine-wire cores (stranded cables) 1 = Red ferrules, ...
Page 23
Proline 500 PROFINET Electrical connection Sensor Electrode cable Coil current cable 70 (2.76) 80 (3.15) 17 (0.67) 40(1.57) 8 (0.31) 8 (0.31) ≥ 1 (0.04) A0029438 A0029439 Engineering unit mm (in) A = Terminate the cable B = Fit ferrules on cables with fine-wire cores (stranded cables) 1 = Red ferrules, ...
Page 24
Electrical connection Proline 500 PROFINET Preparing the connecting cable: Promag P and Promag W Transmitter Electrode cable Coil current cable (3.54) 100 (3.94)* 70 (2.76) 80 (3.15) 50 (1.97) 17 (0.67) 50 (1.97) 8 (0.3) 10 (0.4) 8 (0.3) A0029329...
Electrical connection Proline 500 PROFINET Connecting the measuring device NOTICE Limitation of electrical safety due to incorrect connection! ‣ Have electrical connection work carried out by appropriately trained specialists only. ‣ Observe applicable federal/national installation codes and regulations. ‣ Comply with local workplace safety regulations.
Page 27
Proline 500 PROFINET Electrical connection Proline Promag P and Promag W E2 S2 42 41 36 37 n.c. n.c. n.c. 42 41 E1 E2 A0029145 Coil current cable Signal cable Connecting the connecting cable to the sensor connection housing Connection via terminals with order code for "Housing"...
Page 28
Electrical connection Proline 500 PROFINET Connecting the sensor connection housing via terminals 10 (0.4) 3 mm 22 mm 24 mm A0029612 Loosen the securing clamp of the housing cover. Unscrew the housing cover. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Page 29
Proline 500 PROFINET Electrical connection Connecting the sensor connection housing via terminals 10 (0.4) 8 mm 22 mm 24 mm A0029617 Release the securing screw of the housing cover. Open the housing cover. Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Page 30
Electrical connection Proline 500 PROFINET Connecting the connecting cable to the transmitter 10 (0.4) 3 mm 22 mm 24 mm A0029592 Loosen the securing clamp of the connection compartment cover. Unscrew the connection compartment cover. Push the cable through the cable entry. To ensure tight sealing, do not remove the sealing ring from the cable entry.
Page 31
Proline 500 PROFINET Electrical connection 5.2.2 Connecting the transmitter A0026781 Terminal connection for supply voltage Terminal connection for signal transmission, input/output Terminal connection for signal transmission, input/output or terminal connection for network connection via service interface (CDI-RJ45) Protective earth (PE)
Page 32
Electrical connection Proline 500 PROFINET N ic N ic t e r t e r e r g ö f f e r g ö f f iz e iz e t e n t e n r ir r ir...
Page 33
Proline 500 PROFINET Electrical connection 10. Plug in the RJ45 connector. 11. Firmly tighten the cable glands. This concludes the PROFINET connection process. Connecting the supply voltage and additional inputs/outputs 10 (0.4) mm (in) – A0033983 Push the cable through the cable entry . To ensure tight sealing, do not remove the sealing ring from the cable entry.
Page 34
Electrical connection Proline 500 PROFINET Fit the display module holder in the electronics compartment. Screw on the connection compartment cover. Secure the securing clamp of the connection compartment cover. Removing a cable 3 (0.12) A0029598 3 Engineering unit mm (in)
Page 35
Proline 500 PROFINET Electrical connection 5.2.3 Integrating the transmitter into a network This section only presents the basic options for integrating the device into a network. Integrating via the service interface The device is integrated via the connection to the service interface (CDI-RJ45).
Page 36
Electrical connection Proline 500 PROFINET Integrating into a ring topology The device is integrated via the terminal connection for signal transmission (output 1) and the connection to the service interface (CDI-RJ45). Note the following when connecting: • Recommended cable: CAT5e, CAT6 or CAT7, with shielded connector •...
Proline 500 PROFINET Electrical connection Hardware settings 5.3.1 Setting the device name A measuring point can be quickly identified within a plant on the basis of the tag name. The tag name is equivalent to the device name (name of station of the PROFINET specification).
Page 38
Electrical connection Proline 500 PROFINET Setting the device name Risk of electric shock when opening the transmitter housing. ‣ Before opening the transmitter housing: ‣ Disconnect the device from the power supply. The default IP address may not be activated .
• Grounding rings can be ordered separately as an accessory from Endress+Hauser . When ordering make sure that the grounding rings are compatible with the material used for the...
Page 40
Electrical connection Proline 500 PROFINET Potential equalization via additional grounding ring A0028971 Hexagonal-headed bolts of process connection O-ring seals Plastic disk (spacer) or grounding ring Sensor Potential equalization via grounding electrodes on process connection A0028972 Hexagonal-headed bolts of process connection...
Page 41
Proline 500 PROFINET Electrical connection 5.4.2 Promag P and Promag W For devices intended for use in hazardous locations, please observe the guidelines in the Ex documentation (XA). Metal, grounded pipe A0016315 4 Potential equalization via measuring tube Unlined and ungrounded metal pipe This connection method also applies in situations where: •...
Page 42
Electrical connection Proline 500 PROFINET Plastic pipe or pipe with insulating liner This connection method also applies in situations where: • The customary potential equalization is not used • Equalizing currents are present Ground cable Copper wire, at least 6 mm (0.0093 in...
Proline 500 PROFINET Electrical connection Connect the measuring device to the power supply such that it is floating in relation to the protective ground (isolation transformer). Ensuring the degree of protection The measuring device fulfills all the requirements for the IP66/67 degree of protection, Type 4X enclosure.
Operation options Proline 500 PROFINET Operation options Overview of operation options A0034513 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Mobile handheld terminal Control system (e.g.
Proline 500 PROFINET Operation options Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN 7 Schematic structure of the operating menu 6.2.2...
Operation options Proline 500 PROFINET Access to the operating menu via the local display X X X X X X 19.184 mA 12.5 X X X X X X Language 20.50 Ã English Deutsch Español Français Xxxxxxx A B C D E F G H I J K M N O P Q R S T U V W +0.000 Xx...
Page 47
Proline 500 PROFINET Operation options 6.3.1 Operational display Explanatory symbols for the measured value Status area • Depends on the device version, e.g.: The following symbols appear in the status area of the – : Volume flow operational display at the top right: –...
Page 48
Operation options Proline 500 PROFINET Numeric editor Confirms selection. Moves the input position one position to the left. Exits the input without applying the changes. Inserts decimal separator at the input position. Inserts minus sign at the input position. Clears all entered characters.
Proline 500 PROFINET Operation options Keys and meaning Reduces the contrast (brighter setting). Plus/Enter key combination (press and hold down the keys simultaneously) Increases the contrast (darker setting). Minus/Plus/Enter key combination (press the keys simultaneously) For operational display:Enables or disables the keypad lock.
System integration Proline 500 PROFINET System integration Overview of device description files 7.1.1 Current version data for the device Firmware version 01.00.zz • On the title page of the Operating instructions • On the transmitter nameplate • Firmware version Diagnostics → Device information → Firmware version Release date of firmware version 10.2017...
Page 51
Proline 500 PROFINET System integration 7.2.1 File name of the device master file (GSD) Example of the name of a device master file: GSDML-V2.3.x-EH-PROMAG 500-yyyymmdd.xml GSDML Description language V2.3.x Version of the PROFINET specification Endress+Hauser PROMAG Instrument family Transmitter yyyymmdd Date of issue (yyyy: year, mm: month, dd: day) .xml...
System integration Proline 500 PROFINET Cyclic data transmissionPromag 7.3.1 Overview of the modules The following tables shows which modules are available to the measuring device for cyclic data exchange. Cyclic data exchange is performed with an automation system. Measuring device...
Page 53
Proline 500 PROFINET System integration Selection: device function Slot Device function Status (meaning) Empty pipe detection • 0 (device function not active) 1…10 • 1 (device function active) Low flow cut off Diagnose Input module Transmit discrete input values (diagnostic information) from the measuring device to the automation system.
Page 54
System integration Proline 500 PROFINET Totalizer Control submodule Control the totalizer via the automation system. Selection: control totalizer Slot Sub-slot Value Control totalizer Totalize Reset + hold Preset + hold 11…13 Reset + totalize Preset + totalize Hold Totalizer Mode submodule Configure the totalizer via the automation system.
Page 55
Proline 500 PROFINET System integration Density Temperature Unit code Unit Unit code Unit 1099 kg/dm 1000 1103 kg/l 1003 °R 1097 kg/m 1628 SD4°C 1629 SD15°C 1630 SD20°C 32833 SG4°C 32832 SG15°C 32831 SG20°C 1107 lb/ft 1108 lb/gal (us) 32836 lb/bbl (us;liq.)
Page 56
System integration Proline 500 PROFINET Discrete Output module Transmit discrete output values from the automation system to the measuring device. Assigned device functions Slot Device function Status (meaning) • 0 (disable device function) Flow override • 1 (enable device function) Relay output value: 18…20...
Page 57
Proline 500 PROFINET System integration Verification control Start verification (output data) A change in the status from 0 to 1 starts the verification 7.3.3 Status coding Status Coding (hex) Meaning BAD - Maintenance alarm 0x24 A measured value is not available because a device error has occurred.
Proline 500 PROFINET Commissioning Commissioning Function check Before commissioning the measuring device: ‣ Make sure that the post-installation and post-connection checks have been performed. • "Post-installation check" checklist → 16 • "Post-connection check" checklist → 43 Setting the operating language...
Diagnostic information Proline 500 PROFINET Configuring the measuring device The Setup menu with its submenus and guided wizards is used for fast commissioning of the device. They contain all the parameters required for configuration, such as for measurement or communication.
Page 61
Proline 500 PROFINET Diagnostic information X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2 Diagnostics 3 Supply voltage (ID:203) S801 0d00h02m25s...