Endress+Hauser Proline 400 Brief Operating Instructions
Endress+Hauser Proline 400 Brief Operating Instructions

Endress+Hauser Proline 400 Brief Operating Instructions

Transmitter with ultrasonic time-of-flight sensor modbus rs485
Hide thumbs Also See for Proline 400:

Advertisement

Quick Links

KA01660D/06/EN/01.24-00
71645374
2024-05-15
Products
Brief Operating Instructions
Proline 400
Transmitter with ultrasonic time-of-flight sensor
Modbus RS485
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline 400 and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Proline 400

  • Page 1 Solutions Services KA01660D/06/EN/01.24-00 71645374 2024-05-15 Brief Operating Instructions Proline 400 Transmitter with ultrasonic time-of-flight sensor Modbus RS485 These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 2 of 2: Transmitter Contain information about the transmitter.
  • Page 2 Proline 400 Modbus RS485 Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are Brief Operating Instructions Part 2: Transmitter. The "Brief Operating Instructions Part 1: Sensor" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer •...
  • Page 4: Table Of Contents

    Table of contents Proline 400 Modbus RS485 Table of contents About this document ............. . 5 Symbols used .
  • Page 5: About This Document

    Proline 400 Modbus RS485 About this document About this document Symbols used 1.1.1 Safety symbols DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in serious or fatal injury. WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in serious or fatal injury.
  • Page 6 About this document Proline 400 Modbus RS485 Symbol Meaning Potential equalization connection (PE: protective earth) Ground terminals that must be connected to ground prior to establishing any other connections. The ground terminals are located on the interior and exterior of the device: •...
  • Page 7: Safety Instructions

    Proline 400 Modbus RS485 Safety instructions Safety instructions Requirements for the personnel The personnel must fulfill the following requirements for its tasks: ‣ Trained, qualified specialists must have a relevant qualification for this specific function and task. ‣ Are authorized by the plant owner/operator.
  • Page 8: Workplace Safety

    Safety instructions Proline 400 Modbus RS485 Workplace safety When working on and with the device: ‣ Wear the required personal protective equipment as per national regulations. Operational safety Damage to the device! ‣ Operate the device in proper technical condition and fail-safe condition only.
  • Page 9: Product Description

    Proline 400 Modbus RS485 Product description Product description The measuring system consists of a transmitter and two or one sensor sets. The transmitter and sensor sets are mounted in physically separate locations. They are interconnected by sensor cables. • Prosonic Flow I 400: The sensors function as sound generators and sound receivers. The sensors of a sensor pair are always arranged opposite each other and send/receive the ultrasonic signals directly (single-traverse positioning).
  • Page 10: Special Mounting Instructions

    Mounting procedure Proline 400 Modbus RS485 Unlock the display module. Pull out the display module and turn it to the desired position in increments of 90°. 4.1.2 Mounting the transmitter housing WARNING Excessive tightening torque applied to the fixing screws! Damage to the transmitter.
  • Page 11: Transmitter Post-Installation Check

    Proline 400 Modbus RS485 Mounting procedure Transmitter post-installation check The post-installation check must always be performed after the following tasks: • Turning the transmitter housing • Turning the display module Is the device undamaged (visual inspection)?  Turning the transmitter housing: •...
  • Page 12: Electrical Connection

    Electrical connection Proline 400 Modbus RS485 Electrical connection WARNING Live parts! Incorrect work performed on the electrical connections can result in an electric shock. ‣ Set up a disconnecting device (switch or power-circuit breaker) to easily disconnect the device from the supply voltage.
  • Page 13 Proline 400 Modbus RS485 Electrical connection Connecting cable between the transmitter and sensor Sensor cable for sensor - transmitter A0045277 A0044949 Standard cable • TPE: -40 to +80 °C (-40 to +176 °F) • TPE halogen-free: -40 to +80 °C (-40 to +176 °F) •...
  • Page 14 Electrical connection Proline 400 Modbus RS485 5.2.3 Terminal assignment Transmitter The sensor can be ordered with terminals. Connection methods available Possible options for order code Power "Electrical connection" Outputs supply Terminals Terminals • Option A: coupling M20x1 • Option B: thread M20x1 •...
  • Page 15: Connecting The Measuring Device

    Proline 400 Modbus RS485 Electrical connection Grounding of the cable shield NOTICE In systems without potential matching, the multiple grounding of the cable shield causes mains frequency equalizing currents! Damage to the bus cable shield. ‣ Only ground the bus cable shield to either the local ground or the protective ground at one end.
  • Page 16 Electrical connection Proline 400 Modbus RS485 5.3.1 Connecting the sensor with transmitter WARNING Risk of damaging electronic components! ‣ Connect the sensor and transmitter to the same potential equalization. ‣ Only connect the sensor to a transmitter with the same serial number.
  • Page 17 Proline 400 Modbus RS485 Electrical connection Connect sensor cable to channel 1 downstream. For a two-path measurement: proceed as per steps 3+4 Connect sensor cable to channel 2 upstream. Connect sensor cable to channel 2 downstream. 10. Tighten the cable gland(s).
  • Page 18 Electrical connection Proline 400 Modbus RS485 5.3.2 Connecting the transmitter WARNING Housing degree of protection may be voided due to insufficient sealing of the housing. ‣ Screw in the screw without using any lubricant. The threads on the cover are coated with a dry lubricant.
  • Page 19: Special Connection Instructions

    Proline 400 Modbus RS485 Electrical connection Special connection instructions 5.4.1 Connection examples Modbus RS485 A0028765  3 Connection example for Modbus RS485, non-hazardous area and Zone 2/Div. 2 Control system (e.g. PLC) Ground cable shield at one end. The cable shield must be grounded at both ends to comply with EMC requirements;...
  • Page 20: Post-Connection Check

    Electrical connection Proline 400 Modbus RS485 To ensure that moisture does not enter the cable entry, route the cable so that it loops down before the cable entry ("water trap").  A0029278 The cable glands supplied do not ensure housing protection when not in use. They must therefore be replaced by dummy plus corresponding to the housing protection.
  • Page 21: Operation Options

    Proline 400 Modbus RS485 Operation options Operation options Overview of operation methods A0046501 Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370...
  • Page 22: Structure And Function Of The Operating Menu

    Operation options Proline 400 Modbus RS485 Structure and function of the operating menu 6.2.1 Structure of the operating menu Operating menu for operators and maintenances Language Operation task-oriented Setup Diagnostics Operating menu for experts Expert function-oriented A0014058-EN  4 Schematic structure of the operating menu 6.2.2...
  • Page 23: Access To Operating Menu Via Web Browser

    Proline 400 Modbus RS485 Operation options Access to operating menu via web browser 6.3.1 Function range With the integrated web server, the device can be operated and configured via a web browser service interface (CDI-RJ45) or WLAN interface. The structure of the operating menu is the same as for the local display.
  • Page 24 Operation options Proline 400 Modbus RS485 Computer settings Settings Interface CDI-RJ45 WLAN User rights Appropriate user rights (e.g. administrator rights) for TCP/IP and proxy server settings are necessary (e.g. for adjusting the IP address, subnet mask etc.). Proxy server settings of the web The web browser setting Use a proxy server for your LAN must be disabled .
  • Page 25 Proline 400 Modbus RS485 Operation options Switch on the measuring device. Connect the computer to the RJ45 plug via the standard Ethernet cable . If a 2nd network card is not used, close all the applications on the notebook.  Applications requiring Internet or a network, such as e-mail, SAP applications, Internet or Windows Explorer.
  • Page 26 Operation options Proline 400 Modbus RS485 Enter the password: Serial number of the measuring device ex-works (e.g. L100A802000).  The LED on the display module flashes. It is now possible to operate the measuring device with the web browser, FieldCare or DeviceCare.
  • Page 27 Proline 400 Modbus RS485 Operation options 6.3.5 User interface A0029418 Function row Local display language Navigation area Header The following information appears in the header: • Device name • Device tag • Device status with status signal • Current measured values...
  • Page 28 Operation options Proline 400 Modbus RS485 Functions Meaning Configuration and checking of all the parameters required for establishing the connection to the measuring device: Network • Network settings (e.g. IP address, MAC address) • Device information (e.g. serial number, firmware version)
  • Page 29: Access To The Operating Menu Via The Operating Tool

    Proline 400 Modbus RS485 System integration Enabling the Web server If the Web server is disabled it can only be re-enabled with the Web server functionality parameter via the following operating options: • Via local display • Via Bedientool "FieldCare"...
  • Page 30: Commissioning

    Commissioning Proline 400 Modbus RS485 Commissioning Installation and function check Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • "Post-mounting check" checklist →  11 • "Post-connection check" checklist →  20 Switching on the measuring device ‣...
  • Page 31: Configuring The Measuring Device

    Proline 400 Modbus RS485 Commissioning X X X X X X X 20.50 Main menu 0104-1 Display language English Operation Setup Display language 0104-1 English à Deutsch Español Français Display language 0104-1 à English Deutsch Español Français Hauptmenü 0104-1 Sprache...
  • Page 32: Diagnostic Information

    Diagnostic information Proline 400 Modbus RS485 Submenu Configuration Application Totalizer Diagnosis Diagnostics list, event logbook, device information, simulation Diagnostic information Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. The message about remedial measures can be called up from the diagnostic message, and contains important information on the fault.
  • Page 33 Proline 400 Modbus RS485 Diagnostic information X X X X X X X X X X X X X X 20.50 S801 Supply voltage Menu Diagnostic list Diagnostics 1 S801 Supply voltage Diagnostics 2 Diagnostics 3 Supply voltage (ID:203) S801 0d00h02m25s...
  • Page 36 *71645374* 71645374 www.addresses.endress.com...

Table of Contents