Endress+Hauser Proline 500-digital Brief Operating Instructions
Endress+Hauser Proline 500-digital Brief Operating Instructions

Endress+Hauser Proline 500-digital Brief Operating Instructions

Hart transmitter with ultrasonic time-of-flight sensor

Advertisement

Quick Links

KA01377D/06/EN/02.24-00
71636522
2024-01-31
Products
Brief Operating Instructions
Proline 500 – digital
HART transmitter
with ultrasonic time-of-flight sensor
These instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Brief Operating Instructions part 2 of 2: Transmitter
Contain information about the transmitter.
Brief Operating Instructions part 1 of 2: Sensor →  3
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Proline 500-digital and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Proline 500-digital

  • Page 1 Products Solutions Services KA01377D/06/EN/02.24-00 71636522 2024-01-31 Brief Operating Instructions Proline 500 – digital HART transmitter with ultrasonic time-of-flight sensor These instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Brief Operating Instructions part 2 of 2: Transmitter Contain information about the transmitter.
  • Page 2 Proline 500 – digital HART Order code: XXXXX-XXXXXX Ser. no.: XXXXXXXXXXXX Ext. ord. cd.: XXX.XXXX.XX Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3 These Brief Operating Instructions are Brief Operating Instructions Part 2: Transmitter. The "Brief Operating Instructions Part 1: Sensor" are available via: • Internet: www.endress.com/deviceviewer • Smart phone/tablet: Endress+Hauser Operations App Detailed information about the device can be found in the Operating Instructions and the other documentation: • Internet: www.endress.com/deviceviewer •...
  • Page 4: Table Of Contents

    Diagnostic information ............35 Endress+Hauser...
  • Page 5: About This Document

    Result of a step Visual inspection 1.1.3 Electrical symbols Symbol Meaning Symbol Meaning Direct current Alternating current Direct current and alternating current Ground connection A grounded terminal which, as far as the operator is concerned, is grounded via a grounding system. Endress+Hauser...
  • Page 6 Allen key Open-ended wrench 1.1.6 Symbols in graphics Symbol Meaning Symbol Meaning 1, 2, 3,... Item numbers … Series of steps A, B, C, ... Views A-A, B-B, C-C, ... Sections Hazardous area Safe area (non-hazardous area) Flow direction Endress+Hauser...
  • Page 7: Safety Instructions

    Danger of breakage due to corrosive or abrasive fluids and ambient conditions! ‣ Verify the compatibility of the process fluid with the sensor material. ‣ Ensure the resistance of all fluid-wetted materials in the process. ‣ Keep within the specified pressure and temperature range. Endress+Hauser...
  • Page 8: Workplace Safety

    Verification for borderline cases: ‣ For special fluids and fluids for cleaning, Endress+Hauser is glad to provide assistance in verifying the corrosion resistance of fluid-wetted materials, but does not accept any warranty or liability as minute changes in the temperature, concentration or level of contamination in the process can alter the corrosion resistance properties.
  • Page 9: Device-Specific It Security

    IEC/ISA62443 or the IEEE, is recommended. This includes organizational security measures such as the assignment of access authorization as well as technical measures such as network segmentation. Transmitters with an Ex de approval may not be connected via the service interface (CDI- RJ45)! Endress+Hauser...
  • Page 10: Product Description

    The transmitter and sensor are mounted in physically separate locations. They are interconnected by a connecting cable. Transmitter Connecting cable: cable, separate, standard Sensor connection housing with integrated ISEM (intelligent sensor electronics module) For detailed information on the product description, see the Operating Instructions for the device →  3 Endress+Hauser...
  • Page 11: Mounting Procedure

    Excessive force can damage the housing! ‣ Avoid excessive mechanical stress. The transmitter can be mounted in the following ways: • Wall mounting →  11 • Pipe mounting→  13 4.2.1 Wall mounting Required tools: Drill with drill bit ⌀ 6.0 mm Endress+Hauser...
  • Page 12  1 Engineering unit mm (in) Depends on order code for "Transmitter housing" Order code for "Transmitter housing" • Option A, aluminum, coated: L = 14 mm (0.55 in) • Option D, polycarbonate: L = 13 mm (0.51 in) Endress+Hauser...
  • Page 13 Excessive tightening torque applied to the fixing screws! Risk of damaging the plastic transmitter. ‣ Tighten the fixing screws as per the tightening torque: 2.5 Nm (1.8 lbf ft) ø 20…70 TX 25 ø ( 0.79…2.75) SW 10 A0029051  2 Engineering unit mm (in) Endress+Hauser...
  • Page 14: Transmitter Post-Installation Check

    Mounting the transmitter housing: • Post mounting • Wall mounting Is the device undamaged (visual inspection)?  Post mounting:  Have the fixing screws been tightened with the correct tightening torque? Wall mounting:  Are the securing screws tightened securely? Endress+Hauser...
  • Page 15: Electrical Connection

    Cable diameter • Cable glands supplied: M20 × 1.5 with cable ⌀ 6 to 12 mm (0.24 to 0.47 in) • Spring-loaded terminals: Suitable for strands and strands with ferrules. Conductor cross-section 0.2 to 2.5 mm (24 to 12 AWG). Endress+Hauser...
  • Page 16 Pulse /frequency /switch output Standard installation cable is sufficient Double pulse output Standard installation cable is sufficient Relay output Standard installation cable is sufficient. Current input 0/4 to 20 mA Standard installation cable is sufficient Status input Standard installation cable is sufficient Endress+Hauser...
  • Page 17 Maximum 26 µH IIC, maximum 104 µH IIB Inductance/resistance ratio Maximum 8.9 µH/Ω IIC, maximum 35.6 µH/Ω IIB (e.g. according to IEC 60079-25) (L/R) Loop resistance Power supply line (+, –): maximum 5 Ω Cable length Maximum 150 m (500 ft), see the following table. Endress+Hauser...
  • Page 18 • A, B = 0.5 mm 5.2.4 Terminal assignment Transmitter: supply voltage, input/outputs The terminal assignment of the inputs and outputs depends on the individual order version of the device. The device-specific terminal assignment is documented on an adhesive label in the terminal cover. Endress+Hauser...
  • Page 19 Remove dummy plug if present. If the measuring device is supplied without cable glands: Provide suitable cable gland for corresponding connecting cable. If the measuring device is supplied with cable glands: Observe requirements for connecting cables →  15. Endress+Hauser...
  • Page 20: Connecting The Measuring Device

    Protective earth (PE) Connecting cable ISEM communication Grounding via ground connection; in the version with a device plug, grounding is ensured through the plug itself Cable entry for cable or connection of device plug on sensor connection housing Protective earth (PE) Endress+Hauser...
  • Page 21 Option A "Aluminum, coated" →  22 Prosonic Flow G Option L "Cast, stainless" →  22 Prosonic Flow G Connecting the connecting cable to the transmitter The cable is connected to the transmitter via terminals →  23. Endress+Hauser...
  • Page 22 Housing degree of protection voided due to insufficient sealing of the housing. ‣ Screw in the thread on the cover without using any lubricant. The thread on the cover is coated with a dry lubricant. Screw on the housing cover. Tighten the securing clamp of the housing cover. Endress+Hauser...
  • Page 23  The process for connecting the connecting cable is now complete. Close the housing cover. 10. Tighten the securing screw of the housing cover. 11. After connecting the connecting cable: Connect the signal cable and the supply voltage cable →  24. Endress+Hauser...
  • Page 24 Terminal connection for signal transmission, input/output; optional: connection for external WLAN antenna Protective earth (PE) TX 20 TX 20 10 (0.4) 22 mm 24 mm A0029597 Loosen the 4 fixing screws on the housing cover. Open the housing cover. Fold open the terminal cover. Endress+Hauser...
  • Page 25: Ensuring Potential Equalization

    To guarantee the degree of protection IP66/67, Type 4X enclosure, carry out the following steps after electrical connection: Check that the housing seals are clean and fitted correctly. Dry, clean or replace the seals if necessary. Tighten all housing screws and screw covers. Firmly tighten the cable glands. Endress+Hauser...
  • Page 26: Post-Connection Check

    Are all cable glands installed, securely tightened and leak-tight? Cable run with "water trap" →  25?  Is the terminal assignment correct ?  Are dummy plugs inserted in unused cable entries and have transportation plugs been replaced with  dummy plugs? Endress+Hauser...
  • Page 27: Operation Options

    Local operation via display module Computer with Web browser (e.g. Internet Explorer) or with operating tool (e.g. FieldCare, DeviceCare, AMS Device Manager, SIMATIC PDM) Field Xpert SFX350 or SFX370 Field Xpert SMT70 Mobile handheld terminal Control system (e.g. PLC) Endress+Hauser...
  • Page 28: Structure And Function Of The Operating Menu

    The individual parts of the operating menu are assigned to certain user roles (e.g. operator, maintenance etc.). Each user role contains typical tasks within the device life cycle. For detailed information on the operating philosophy, see the Operating Instructions for the device. →  3 Endress+Hauser...
  • Page 29: Access To Operating Menu Via Local Display

    2.6 Explanatory symbols for measured value 2: measured value type, measuring channel number Navigation view: picklist of a parameter 3.1 Navigation path and status area 3.2 Display area for navigation:  designates the current parameter value Editing view: text editor with input mask Editing view: numeric editor with input mask Endress+Hauser...
  • Page 30 Moves the input position one position to the left. Switches to the selection of the correction tools. Deletes one character immediately to the left of the input position. Toggle • Between upper-case and lower-case letters • For entering numbers • For entering special characters Endress+Hauser...
  • Page 31 If present, opens the help text for the function of the parameter. In wizards Opens the editing view of the parameter and confirms the parameter value In the text and numeric editor • Pressing the key briefly confirms your selection. • Pressing the key for 2 s confirms your entry. Endress+Hauser...
  • Page 32: Access To The Operating Menu Via The Operating Tool

    For detailed information on access via FieldCare and DeviceCare, see the Operating Instructions for the device →  3 Access to the operating menu via the web server The operating menu can also be accessed via the web server. See the Operating Instructions for the device.→  3 Endress+Hauser...
  • Page 33: System Integration

    Before commissioning the device: ‣ Make sure that the post-installation and post-connection checks have been performed successfully. • "Post-mounting check" checklist →  14 • "Post-connection check" checklist →  26 Setting the operating language Factory setting: English or ordered local language Endress+Hauser...
  • Page 34: Configuring The Measuring Device

    System units Configuration of the units for all measured values I/O configuration User configurable I/O module Current input Configuration of the input/output type Status input Current output 1 to n Pulse/frequency/switch output 1 to n Relay output Double pulse output Endress+Hauser...
  • Page 35: Protecting Settings From Unauthorized Access

    Faults detected by the self-monitoring system of the measuring device are displayed as a diagnostic message in alternation with the operational display. The message about remedial measures can be called up from the diagnostic message, and contains important information on the fault. Endress+Hauser...
  • Page 36  The Diagnostic list submenu opens. Select the desired diagnostic event with  or  and press  .  The message about the remedial measures opens. Press  +  simultaneously.  The message about the remedial measures closes. Endress+Hauser...
  • Page 40 *71636522* 71636522 www.addresses.endress.com...

Table of Contents