Emerson Rosemount 2120 Quick Start Manual page 11

Level switch
Hide thumbs Also See for Rosemount 2120:

Advertisement

June 2020
3.5
Wiring diagrams
CAUTION
Before use, check the cable glands and blanking plugs are suitably rated.
Isolate supply before connecting the switch or removing the electronics.
The Protective Earth (PE) terminal must be connected to an external
earthing system.
3.5.1
Direct load switching electronics cassette
Figure 3-1: Direct Load Switching (Two-wire, Red Label)
U
=
I
<
OFF
I
=
L
I
=
PK
Quick Start Guide
1 2 3
20 - 264 V ~ (ac) (50/60 Hz)
4 mA
20 - 500 mA
5 A, 40 ms (inrush)
Isolate supply before making connections.
Example of external wiring
1
R
PE
(Ground)
Neutral
0 V
R
=
External load (must be fitted)
U
=
20 - 60 V
I
<
4 mA
OFF
I
=
20 - 500 mA
L
I
=
5 A, 40 ms (inrush)
PK
Quick Start Guide
2
3
Fuse
2A(T)
DPST
I L
Live
+ V
(dc)
11