Invacare® Birdie® EVO
Symptoms
The control unit emits a beeping sound
during lifting, and the motor stops.
Manual legs fail to open with the pedal
or lever.
Damping of the spreader bar has
lessened or disappeared completely.
The detachable spreader bar does not
fix to the connector on the boom
Contact your Invacare provider or representative if the above does not solve your problems.
10
Faults
Max. load is exceeded
The mechanical mechanism is damaged Replace foot pedal and/or rods. See
Damper is damaged
The locking mechanism is damaged
Solution
Reduce the load (and the lift will
function normally).
5.14 Replacing the foot pedal and/or
rods, page 22
Replace the damper. See 5.9 Replacing
the damper, page 16
Replace the spreader bar connector.
See 5.9 Replacing the damper, page 16