Main unit specifications
Analog
Measurement
Resistance
input
accuracy
temperature
section
(*2) (*3)
detector
Maximum input voltage
Reference contact compensation
Input impedance
Allowable signal source resistance
Temperature coefficient
Withstand voltage
Insulation resistance
External
Logic input/
Number of channels
Input/
Pulse input
(*4)
Output
Pulse
Sections
input
modes
Trigger input
Number of channels
Specifications
Maximum input
of each input
voltage
section
Threshold voltage
Hysteresis range
Alarm output
Number of channels
Output format
Output conditions
ZR-RX40 Start Up Guide
ZR-RX40A
Type
Measurement
Temperature Range (° C)
–200 to 850 (FS = 1050 ° C) 1 mA
Pt100
–200 to 500 (FS = 700 ° C) 1 mA
JPt100
–200 to 500 (FS = 700 ° C) 0.2 mA
Pt1000
60 Vp-p (between +/– terminals, between input terminals, between
input terminal/GND)
Internal/External switching
1 MΩ ±5 %
300 Ω or less
Gain: 0.01 % of F.S./° C
350 Vp-p (between input channel/GND; between each channel) 1
minute
Between input terminal/GND: At least 50 MΩ (at 500 VDC)
4 ch
Revolutions
Spans
mode
Maximum number of pulse
inputs
Counts
Spans
mode
Maximum number of pulse
inputs
Inst. Mode Spans
Maximum number of pulse
inputs
1 ch
24 V
Approx. 2.5 V
Approx. 0.5 V (2.5 to 3 V)
4 ch
Open collector output (pull-up resistance: 10 kΩ)
5 to 24 VDC, up to 100 mA
Level judgment, window judgment, logic pattern judgment, pulse
judgment
Applied current
Measurement
Accuracy
±1.0 ° C
±0.8 ° C
±0.8 ° C
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M RPM/F.S.
50 k/sec
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M C/F.S.
50 k/sampling interval
50, 500, 5000, 50 k, 500 k, 5 M, 50 M,
500 M C/F.S.
50 k/sampling interval
Standard Specifications
43