Forced Air and Convection

Since the Product has a large surface area, a major component of the temperature uncertainty is
caused by changes in convection. Because of this, keep the Product away from areas with large
amounts of air flow or drafts.
Table 5. Apparent Temperature Limits
4180
HI (°C)
LO (°C)
120.0
-15.0
120.0
-15.0
120.0
-15.0
120.0
-15.0
120.0
-15.0
120.0
-15.0
119.2
-14.5
118.4
-14.0
117.6
-13.5
116.8
-13.0
116.0
-12.5
Precision Infrared Calibrator
4181
HI (°C)
LO (°C)
500.0
35.0
500.0
35.0
500.0
35.0
500.0
35.0
500.0
35.0
500.0
35.0
496.6
35.0
493.2
35.0
489.8
35.0
486.4
35.0
483.0
35.0