Figure 21 Inputs Screen - Kahn Optidew Series User Manual

Chilled mirror hygrometer
Table of Contents

Advertisement

Optidew User's Manual
Inputs
Source (Pressure
Input)
Pressure Unit
Value
(If 'Fixed'
selected)
Range Low
(If 'External'
selected)
Range High
(If 'External'
selected)
Compensation
Direction (If
'Compensation'
On)
Source
(Temperature
Input)
Value
(If 'Fixed'
selected)
Kahn Instruments
Figure 21
Inputs screen
Changes between pressure input from external 4-20mA
transmitter or a fixed value
Available input: Fixed, External
Measurement unit for pressure inputs
Available input: kPa, psig, psia, barg, bara
Sets pressure used for internal calculations
Sets the low range of the connected pressure transmitter
Sets the high range of the connected pressure transmitter.
Recalculate dewpoint based on pressure input
Available input: Off, On
Select 'From Atmos' if dewpoint sensor is at atmospheric
pressure.
Select 'To Atmos' if dewpoint sensor is at entered fixed pressure
or pressure measured by transducer.
Changes between temperature input from external PT100 or a
fixed value.
Available input: Fixed, External
Sets temperature used for internal calculations
OPERATION
27