Setup Menus; Figure 15 Dcc Menu - Kahn Optidew Series User Manual

Chilled mirror hygrometer
Table of Contents

Advertisement

Optidew User's Manual
3.2

Setup Menus

DCC (Dynamic Contamination Correction)
Type
Setpoint
Mode
Interval
Period
Output hold
FAST
FAST SetP
Kahn Instruments
Figure 15
DCC heating temperature can either be relative to last measured
dewpoint or an absolute temperature. Actual temperature or Δ is
defined by 'Setpoint'.
Available input: Relative, Absolute
Mirror heating temperature during DCC, either absolute or
relative to last measured dewpoint. See 'Type' option above.
Available input: 1 to 120°C
DCCs can either be triggered automatically at every Interval, or
they can be manually triggered only.
Available input: Manual, Auto
Time between automatic DCCs
Input format: hh:mm
Limits: 01:00 to 99:00
Duration of a DCC
Input format: hh:mm
Limits: 00:01 to 00:59
Minimum time to hold analog outputs after finishing a DCC
Input format: hh:mm
Limits: 00:04 to 00:59
Turns frost assurance on or off. See section 4.4.3 for further
information
Available input: On, Off
Passing this mirror temperature will trigger the frost assurance
function without a DCC
Available input: -28 to -2°C
DCC menu
OPERATION
21