ii) Insert Safety Clamp and pressure disc into retainer.
iii) Ensure infusion set is fully inserted into tubing guide and all air is removed
iv) Use minimal stretching of the silicone segment when loading or priming
7.
Close door and open roller clamp. Ensure no drops are falling in the drip chamber. If
drops are falling in the drip chamber then clamp the infusion set immediately, open
door and verify correct loading of the infusion set.
0000CF04424.indd 5
from the infusion set.
the infusion set.
Pushing the Safety Clamp Slider completely into Safety Clamp Frame
may lead to uncontrolled flow to the patient. Therefore, always close
the roller clamp before pushing in the Safety Clamp Slider, if required.
Safety Clamp
Tubing Guide
0000CF04424 Iss. 1 5/18