Display
BATTERY EMPTY
The internal battery is exhausted. The
Pump will automatically switch off in
the immediate future.
SAFETY CLAMP
Safety clamp broken or missing.
SET MISLOAD
Set loaded incorrectly.
FLOW SENSOR DISCONNECT
Flow sensor unplugged during infusion.
WRONG SET
Safety clamp not detected
DOOR CLOSE INCOMPLETE
Safety clamp in non-occluded position
with door open or obstructed.
LEVER OPEN
Door lever is open
VTBI DONE
Intended VTBI completed. When
STOP selected as the END RATE.
ATTENTION
Pump left on hold for 2 minutes after
first ATTENTION warning.
0000CF04424.indd 14
Action
• Switch Pump off and connect to power
supply immediately.
• Clamp infusion set using roller clamp.
• Replace infusion set.
• Investigate and correct set loading.
• Clamp infusion set using roller clamp.
• Investigate and correct set loading.
• Check / replace flow sensor or set VTBI.
• Clamp infusion set using roller clamp.
• Check set and close door.
• Replace infusion set as required.
• Clamp infusion set using roller clamp.
• Investigate and correct set loading.
• Close door.
• Check door lever.
• Check lever hooks.
• Check lever is not obstructed, if so, free
obstruction.
• Set new VTBI or clear VTBI.
• Review Pump setup. Start infusion or turn
off Pump.