BD Alaris VP Plus Guardrails Quick Reference Manual page 15

Volumetric pump
Hide thumbs Also See for Alaris VP Plus Guardrails:

Advertisement

Display
BATTERY LOW
Low battery threshold sensed; remaining
battery run time is limited; if cancelled will
reoccur every ten minutes if not plugged in.
AC POWER FAIL
AC power disconnected or failed.
VTBI DONE
Intended VTBI completed. When KVO or
CONTINUE are selected as the END RATE.
AIR-IN-LINE
Air detected in infusion set at the start of
infusion. Set not fitted correctly into air in
line detector.
SET CLOCK
Date/time not set.
TITRATION
Rate titration not confirmed.
RATE LOCK
Rate lock not confirmed.
LOG FAILURE
Unable to update event log.
SET SERIAL NUMBER
Serial number not set.
NEAR END OF INFUSION
The Pump is nearing the end of the
infusion. This value can be configured.
KVO
Occurs after a VTBI DONE warning is
cancelled if KVO was selected as the END
RATE setting.
0000CF04424.indd 15
Action
• Connect to power supply.
• Check power cable.
• Reconnect to power supply.
• Set new VTBI or clear VTBI.
• Ensure set is fitted correctly in the air
in line detector.
• Assess air in infusion set.
• Check fluid level in drip chamber.
• Check fluid level in container.
• Qualified service personnel must set
date/time.
• Press CANCEL softkey to continue.
• Confirm or cancel new rate.
• Select YES or NO as required.
• Qualified service personnel may need
to service the Pump.
• Contact qualified service personnel
to set the serial number.
• Prepare next infusion, if required.
• Set new infusion or stop treatment
0000CF04424 Iss. 1 15/18