Troubleshooting
#Resource Mgr
#09/09/97 08:48:49
Auto Exit - Off
Identify Static Configure Devices
Found a device at LA 1
Found a device at LA 2
Identify Dynamic Configure Devices
Matching Devices to Slots
match la=1 to slot=1
match la=2 to slot=3
Setting VISA Attributes
la 1, slot 1: device_class 2, manf_id 0xffd, model_code 0x7f4, addr_spc 0
la 2, slot 3: device_class 2, manf_id 0xffd, model_code 0x7f1, addr_spc 0
Setting VISA Address Maps
A24 device @ la 1 - starting address 200000x, size 65536
A24 device @ la 2 - starting address 210000x, size 65536
Enabling Events & Responses
la 1: Int ID 1 assigned to IRQ 4
Enabling Events: 16-32 124-125 127
la 1: Asynchronous Enable succeeded
**Responses are unsupported by this device
la 2: Int ID 1 assigned to IRQ 4
Enabling Events: 16-32 124-125 127
la 2: Asynchronous Enable succeeded
**Responses are unsupported by this device
Begin Normal Operation
slot 1, LA 1, started successfully
slot 3, LA 2, started successfully
VISA Data
la_1=1,1,4093,2036,2,0,1,7,2097152
la_2=2,3,4093,2033,2,0,1,7,2162688
Figure 6–18: ResMan32 program output
6–50
The ResMan32 will check all of the installed modules and their address
locations. The program will print out data similar to that in Figure 6–18. In
this example the portable mainframe has two logic analyzer modules
installed. Both modules were properly configured and the resources were
successfully configured.
If ResMan32 encounters any errors (such as an unsupported instrument or
application module), the resource manager will stop further communications
and display information on why or at what point the instrument module was
disabled.