Monitoring / Preventive And Corrective Maintenance; Monitoring; Preventive And Corrective Maintenance - Satelec ACteon i-Surge+ User Manual

Implant/surgery system

Advertisement

Available languages

Available languages

- Sterilization temperature: 134°C (273°F)//2.1-2.2 bars.
- Time at sterilization temperature: 5 minutes
When treating patients who may have an acute, critical infectious disease, be sure to observe
the hygienic measures cited in applicable publications and reports. If possible, use suitable
disposable products to avoid the transmission of critical pathogens. These protect the user, the patient
and all participants in the surgery. A microbiologically appropriate coolant liquid must be used. Only
use prescribed isotonic saline solution NaCl 0.9 (identified as an infusion solution; follow the
instructions on the package insert) for cooling and spraying wounds.
C. Not all autoclaves draw a pre-vacuum. Please refer to your autoclave manufacturer for specific
sterilization instructions.
XI – MONITORING / PREVENTIVE AND CORRECTIVE MAINTENANCE
Important:
In the case of a fault, you are advised to contact the supplier of your device. Do not use
unauthorized repairers, who might make your device dangerous for you and your patients.

11.1 MONITORING

Regular monitoring of the device and its accessories is necessary in order to detect any insulation fault
or damage. Replace, if necessary.
It is important to keep the control unit ventilation vents clean in order to avoid abnormal heating.

11.2 PREVENTIVE AND CORRECTIVE MAINTENANCE

I-SURGE + micro-motor inspection/overhaul: The manufacturer recommends inspection or overhaul of
the I-SURGE + micro-motor at least once a year.
The micro motor cable has limited durability and is considered a consumable product. The durability of
the cable is largely dependent upon use and treatment, i.e., handling, reprocessing and frequency of
use.
Caution must be exercised when connecting or disconnecting the micro motor and foot control cable
from the device to avoid damage to the cable (Fig. 8). Motor assembly has been tested for 500
sterilization cycles.
Periodically inspect the micro motor cord for damage prior to use. If there is damage noted, contact the
Manufacturer.
1. Do not attempt to disassemble the motor or motor connector.
2. Do not oil or lubricate the motor.
Not all autoclaves can reach 134°
22