3. Do not attach a handpiece to the motor while the motor is running.
4. Do not bend motor cord sharply.
5. Do not drop or impact micro motor against hard surface.
Failure to comply with any of the above instructions may avoid your warranty
The average life of this medical device with normal use and in accord with the present user
manual, has been calculated to be 5 years. According with EN62353 we recommend users
verify every 2 years from the date of the first use: dispersion current versus ground, in the box and in
the patient.
MOTOR RUBBER O-RING:
If necessary, substitute the "Rubber O-Ring".
-
LUBRICATING:
Do not oil or lubricate the micro motor. Protect motor from excess oil draining from
-
handpiece. After lubricating and before autoclaving, stand handpiece by its base on a paper
towel and allow excess oil to drain.
The internal gears of the peristaltic pump are self lubricating.
-
Do apply some oil drops in the cylinders of the peristaltic pump rotor after a long period of
-
non-use or when the pump doesn't run continuously.

11.3 FUSE REPLACEMENT

Your device is protected by two fuses located in the power receptacle. The replacement procedure is as
follows (Fig. 11):
ONCE UNIT IS UNPLUGGED -With a screwdriver remove the fuse holder from the power inlet
-
connector.
Extract the fuse holder, replace the fuse with one of the same type and value 2 X 1A T(230V)
-
/ 2 X 2A T(115V) (Dim. 5x20).
After inspection or replacement of a fuse, re-secure fuse holder ensure that the correct
-
voltage value appears.
Note:
All the information necessary for the repair of faulty components is made available by the manufacturer
to the technical personnel of SATELEC approved dealer network who carry out the work.
 The I-SURGE + must be used in accordance with these instructions for use.
 The I-SURGE + has no components which can be repaired by the user. Assembly,
modifications or repairs must only be undertaken by an authorized service organization.
 Unauthorized opening of the equipment invalidates all claims under warranty and any
other claims.
 Use on electrical circuit complying with the terms of regulation CEI 64-8 section 710.

11.4 OPERATION FAULTS

Fault detected: The central unit does not switch on.
Possible cause
Solution
23