Maintenance - Satelec X-Mind DC Installation Manual

Table of Contents

Advertisement

To guarantee the X-ray system's safety, the following checks must be carried out:
CAUTION
The owner is responsible for arranging and respecting the maintenance schedule.
The X-ray system must undergo maintenance to check for correct operation every 12
months.
1.
Cut the power
2.
Release the spring of the "ARM B" of the pantograph type arm using the key provided
(refer to § 5.7 "BALANCING THE PANTOGRAPH-TYPE ARM")
3.
Remove the tube head
(refer to § 5.6 "ASSEMBLING THE TUBE HEAD")
4.
Remove the wall plate guard
(refer to § 5.2 "ASSEMBLING THE WALL PLATE")
5.
Remove the terminal board cover and disconnect the pantograph-type arm cable
6.
Remove the bracket plug and remove the guard slab
(refer to § 5.4 "ASSEMBLING THE PANTOGRAPH-TYPE ARM")
7.
Remove the pantograph-type arm and its cable from the bracket
8.
Remove the bracket from the wall plate
(refer to § 5.3 "ASSEMBLING THE BRACKET")
9.
Check the wall plate's vertical alignment – adjust if required
10.
Check the six wall plate's adjusting screws – tighten if required
11.
Clean the old lubricating grease from the bracket shaft
12.
Should the bracket shaft be damaged, install a new bracket
13.
Clean the old lubricating grease from the bracket bush
14.
If the bracket bush is damaged, install a new bracket
15.
Grease the bracket shaft (use "molikote D" lubricant)
16.
Lubricate the bracket bush with lubrication grease (use "molikote D" lubricant)
17.
Install the bracket in the wall plate
(refer to § 5.3 "ASSEMBLING THE BRACKET")
18.
Check the pantograph-type arm cable
If it is be damaged, send the pantograph-type arm to the manufacturer for repairs
19.
Check the pantograph-type arm guards
20.
Replace the damaged guards
21.
Remove all old grease of the shaft
(refer to §. 5.4 "ASSEMBLING THE PANTOGRAPH-TYPE ARM")
22.
If the shaft is damaged, send the pantograph-type arm to the manufacturer for repairs
23.
Lubricate the pantograph-type arm shaft with lubricant (use "molikote D" lubricant) and
reposition it in the bracket
24.
Replace the pantograph-type arm cable in the bracket and the wall plate, connect it to the
terminal board and replace the terminal board cover
(refer to § 5.5 "CONNECTION TO THE TERMINAL BOARD")
25.
Position the guard slab in the bracket
26.
Position the bracket plug
27.
Position the plate guard
28.
Check the tubehead's electrical contact
29.
If damaged, send the tube head to the manufacturer for repairs
30.
Clean the old grease from the tube head assembly shaft
31.
Lubricate the tubehead's assembly shaft with a thin layer of lubricant(use "molikote D"
lubricant)
32.
Reposition the tubehead
33.
Load the spring of the pantograph-type arm "ARM B" using the key provided
(refer to § 5.7 "BALANCING THE PANTOGRAPH-TYPE ARM")
34.
Power up and check that the X-ray system is operating correctly
4
4
8
8