•Power cord not connected properly
•Unit has not been switched on.
•Wrong voltage.
•Broken connections inside.
• Transformer damaged.
• PCB faulty.
• Filter supply defective.
• Power interrupted.
Fault detected: The central unit turns off and stops working suddenly
Possible cause
• Power cable damaged.
• Filter damaged.
• Transformer damaged.
• PCB damaged.
• Connections inside interrupted.
Fault detected: The central unit doesn't switch off
Possible cause
•Power filter damaged.
Fault detected: The foot pedal doesn't start the motor
Possible cause
• Micro Motor cable not connected
properly
• Footswitch cable is not connected
properly
•Male connector of the Micro motor
damaged.
•Pins of the connector damaged
•Wrong voltage.
•Foot pedal damaged.
Fault detected: It is not possible to change "Program", FWD/REV and Flow by using the footswitch.
Possible cause
• Footswitch damaged.
•Metal pins damaged in the connector
Fault detected: The display is not working properly
Possible cause
•Contra-angle damaged.
Make sure the power cord is properly plugged into the electrical outlet and
the connector on the back of the unit.
Check the back of the unit to see that the On button has been pressed.
Verify that the voltage matches the power supply. If it does, try to switch
the central unit off and then on again, or connect the device to another
working electrical outlet. If the unit does not turn on, verify the integrity of
the fuses in the unit.
Immediately discontinue use of the device and contact SATELEC after-sales
service.
Solution
After checking the correct connection to the electrical outlet and the
absence of liquid in the unit, try to power the central unit. If the device
does not turn on, immediately discontinue use of the device and contact
SATELEC after-sales service.
Solution
Unplug the power cord from the electrical outlet and the central unit. Then
contact SATELEC after-sales service.
Solution
Check to see that the Micro Motor cable is securely plugged into the
connector on the front right of the central unit, disconnect and reconnect.
Check that the footswitch cord is plugged into the connector on the front
left of the central unit, disconnect and reconnect.
Unplug the Micro Motor and make sure that the male connector is not
damaged, wet or damp.
Unplug the Micro Motor and the footswitch cable, make sure the metal pins
on the male connector are intact and not damaged and have not broken off
and/or remained stuck in the female connector mounted on the unit.
Verify that the voltage matches the power supply. If it does, try to switch
the central unit off and then on again, or connect the device to another
electrical outlet.
Check the ability to start the Micro motor from the central unit keyboard,
pressing the [Motor ON/OFF switch]. If so, contact SATELEC after-sales
service.
Solution
Check the possibility of changing the parameters by using the unit
keyboard.
Contact SATELEC after-sales service.
Solution
Check the integrity of the Contra-angle.
24