Q400/Q500 Only - Removal Of Castor Filler Base - Sunrise Medical Quickie Q400 F Service Instructions Manual

Hide thumbs Also See for Quickie Q400 F:
Table of Contents

Advertisement

Q400/Q500 Only - Removal of castor filler base
Preparation
Note! Switch off the wheelchair via remote control
Instructions
Note! Suitable work area required!
• Follow the illustrated instruction for removing the castor
filler base.
• Rebuild the wheelchair in reverse order once completed.
• In order to complete this, you must first remove the castor
fork.
Fig. 6.2.44
1. Unscrew and remove 3 (A) bolts for underneath the castor arm
Tightening torque for (A) bolts is 7.5Nm
Fig. 6.2.45
2. Slide and remove castor filler base (B)
NOTE: Please note the orientation of the castor filler base
Q400/Q500/Q700 F/R Rev.A
Tools used
5
2
B
Fig. 6.2.44
1
A
Fig. 6.2.45
33

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents