Sunrise Medical Quickie Q400 F Service Instructions Manual page 29

Hide thumbs Also See for Quickie Q400 F:
Table of Contents

Advertisement

Fig. 6.2.35
4. Remove brake (D) from the motor assembly
NOTE: For the standard 2-pole motor (once the brake lever and main
brake cover have been removed):
1. Disconnect electrical connector
2. Unscrew and remove 2 screws through the access holes in brake
using the 3mm Hex Key
3. Remove brake from the motor assembly
NOTE: If finding your old brake doesn't have the connector but is
hardwired then take the following steps:
1. Cut cable to remove old brake leaving enough to reconnect new
brake.
2. Either add new connector to cables from existing motor or
remove connector from new brake and solder wires from existing
motor.
3. Cover soldered joint with heat shrink using heat gun.
Q400/Q500/Q700 F/R Rev.A
Fig. 6.2.35
4
D
29