Waveform: M1, M2, M3, M4; (Store full-length waveforms with 16 bits/data point)
Zoom and Math: Four traces A, B, C, D with chained trace capability
3.15

Setup Storage

For front panel and instrument status: Four non-volatile memories and floppy drive are
standard; hard drive and memory card are optional
3.16
Interface
Remote Control: Full control of all front panel controls and internal functions via GPIB and
RS-232-C, or Ethernet (optional)
GPIB Port: Full control via IEEE-488.2; configurable as talker/listener for computer control
and data transfer
RS-232-C: Asynchronous transfer rate of up to 115.2 kbaud
Ethernet (optional): 10 Base-T Ethernet interface
Floppy Drive: Internal, DOS-format, 3.5" high-density
PC Card Slot (optional): Supports memory and hard drive cards
External Monitor Port Standard: 15-pin D-Type VGA-compatible
Centronics Port: Parallel printer interface
Internal graphics printer (optional): Provides hardcopy output in <10 seconds
3.17
Outputs
Calibrator signal: 500 Hz–1 MHz square wave or DC level, Select from −1.0 to +1.0 into
1MΩ, output on front panel test point and ground lug
Control signals: Rear panel, TTL level, BNC output, Choice of trigger ready, trigger out, or
Pass/Fail status (output resistance 300 Ω ± 10 %)
3.18

Environmental and Safety

Operating Conditions: Temperature 5–40° C rated accuracy, 0-45°C operating, -20° -
60°C non-operating; humidity 80 % RH non-condensing up to 35° C, derates to 50% max
RH, non-condensing at 45°C; 4500 m (15000ft) max up to 25°C; Derates to 2000 m
(6600ft) at 45°C.
CE Approved: EMC Directive 89/336/EEC; EN61326-1 Emissions and immunity, Low
Voltage Directive 73/23/EEC; EN 61010-1 Product Safety (Installation Category II, Pollution
Degree 2)
UL and cUL: UL Standard UL 3111-1; cUL Standard CSA C22.2 No. 1010-1