5.5

Leakage Current

The leakage current is tested by measuring the voltage across the input channel.
Test limit
DC 1MΩ : ±1 mV
5.5.1
Channel Leakage Current
! Recall xxxP010.PNL or configure the DSO:
! Set the DMM to measure Volts, and connect it to Channel 1.
! Measure the voltage and enter it in Table 3. Compare it to the limits.
! Repeat the test for all input channels.
5-8 Performance Verification
Panel Setups
:
Channels Trace ON
Input Coupling
:
Input gain
:
Trigger mode
:
Time base
:
Recall FROM DEFAULT SETUP
Channel 1, Channel 2, Channel 3 & Channel 4
DC 1MΩ Ω Ω Ω on all 4 Channels
50 mV/div. on all 4 Channels
Auto
10 µ µ µ µ sec/div.