5
Menu item
Output current
during error message
Output current
during HOLD
Output current for
HOLD FIX
Output Signal During HOLD:
Output current
[mA]
21
4
Action
Select ON or OFF using
 keys.
Press enter to confirm.
LAST: During HOLD the
last measured value is
maintained at the output.
FIX: During HOLD a value
(to be entered) is main-
tained at the output.
Select using 
Press enter to confirm.
Only with FIX selected:
Enter current which is to
flow at the output during
HOLD
Enter value using 
 keys.
Press enter to confirm.
Output signal for HOLD
FIX setting = 21.0 mA
HOLD active
Configuration
Choices
ON/OFF
LAST/FIX
04.00...22.00 mA
(21.00 mA)
Output signal for HOLD
LAST setting
HOLD active
47