8 Maintenance
A
B
62
Operating Instructions Midwest Stylus
➢ Remove the LED and re-insert it after rotating it 180° around its own
axis.
8.2.4
Replacing the spray water cartridge
Required accessories
● Spray water cartridge
● Midwest Plus Aerosol Spray
If the spray water cartridge leaks or is clogged.
1. Remove the quick coupling from the supply hose.
2. Remove the sealing washer (B).
3. Click the control ring into the maximum water flow. The spray water
C
cartridge (A) can be replaced only in this position.
4. Remove the spray water cartridge.
5. Spray oil lightly onto the new spray water cartridge.
6. Insert the new spray water cartridge.
®
®
Plus / Midwest Tradition
TC/Pro TCF
66 07 290 D3673.201.01.04.02