Dentsply Sirona Midwest Stylus Plus Operating Instructions Manual page 41

Air-driven-high-speed handpiece for connection
Hide thumbs Also See for Midwest Stylus Plus:
Table of Contents

Advertisement

Available languages

Available languages

6607290.book Seite 39 Donnerstag, 4. Januar 2018 11:55 11
®
®
Operating Instructions Midwest Stylus
Plus / Midwest Tradition
TC/Pro TCF
5 Operation
5.4
Adjusting the cooling spray
➢ Adjust the flow rate of the cooling water using the control ring (A) (>
> 50 ml/min
A
50 ml/min).
Tip: You can measure the amount of cooling water with a measuring
cup and watch.
Water flow rate
Quick coupling F (Dentsply Sirona): The maximum water flow is set
Quick coupling F
when the control ring clicks into place.
Multiflex LUX quick coupling (KaVo): The maximum water flow is set
KaVo quick coupling
when the two markings face each other.
MachLite quick coupling (NSK): The maximum water flow is set when
NSK MachLite quick coupling
the control ring clicks audibly into place.
39