Discharging the P-725

The P-725 must be discharged in the following cases:
When the P-725 is not used but the electronics remain switched on to ensure
temperature stability
Before demounting (e.g., before cleaning and transporting the P-725) and for
modifications
If the connecting cable of the P-725 is accidentally pulled out of the electronics during
operation
Requirements
 You have read and understood the safety precautions (p. 6).
Discharging a P-725 Connected to the Electronics
In closed-loop operation:
1. Switch off the servo mode on the electronics.
2. Set the piezo voltage to 0 V on the electronics.
In open-loop operation:
 Set the piezo voltage to 0 V on the electronics.
Discharging a P-725 not Connected to the Electronics
 Connect the P-725's voltage plug to the switched-off electronics from PI for 10 seconds.

Maintenance

NOTICE
Misalignment from loosening screws!
The P-725 is maintenance-free and precision adjusted.
 Loosen screws only when instructed in this user manual.
 Do not open the P-725.