Pifoc long-travel objective scanner, pifoc piezo nanofocusing system for long travel ranges and fast step-and-settle, pifoc high dynamics scanner (40 pages)
Photographic illustrations may also differ and must not be seen as guaranteed properties. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 4 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
8. Click the desired manual and save it to the hard disk of your PC or to a data storage medium. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 5 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Note the depth of the mounting holes in the motion platform (p. 21). Only use screws of the correct length for the respective mounting holes. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 8 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Product View of E-709 Figure 2: Example view of E-709 (see the E-709 user manual for details) Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 10 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
You have read and understood the safety precautions (p. 7). The P-736.ZRx is not connected to the controller. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 12 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Page 13
4. Make sure that the contact resistance at all connection points relevant for mounting the protective earth conductor is <0.1 Ω at 25 A. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 13 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
3. Tighten each of the four screws with a torque between 0.8 Nm and 1.1 Nm. 4. Check that the P-736.ZRx is affixed firmly to the surface. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 14 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Mounting the P-736.ZRx onto an M-545 stage Mount the P-736.ZRx onto the M-545 stage as depicted in the figure. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 15 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
3. Mount the holder with the M2 screws to the provided M2 threaded holes (see Figure 6 above). Pay attention to the specified torque range (p. 22) while doing so. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 16 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Do not do any ultrasonic cleaning. Customer Service For inquiries and orders, contact your PI sales engineer or send us an email (info@pi.ws). If you have questions concerning your system, have the following information ready: Product and serial numbers of all products in the system ...
IEC 60529 *Specifications assured from 17 °C to 23 °C, performance may be reduced outside this range. Physik Instrumente (PI) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany Page 20 / 25 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.ws,...
Dispose of your old equipment according to international, national, and local rules and regulations. In order to fulfil its responsibility as the product manufacturer, Physik Instrumente (PI) GmbH & Co. KG undertakes environmentally correct disposal of all old PI equipment made available on the market after 13 August 2005 without charge.
Need help?
Do you have a question about the P-736.ZRxS PInano and is the answer not in the manual?
Questions and answers