Mounting the P-725 using the Mounting Holes in the Base Body
(Optional)
NOTICE
Warping of the P-725 when mounted on uneven surfaces!
Mounting the P-725 on an uneven surface could warp the P-725. Warping reduces the
accuracy.
 Mount the P-725 onto a flat surface. The recommended flatness of the surface is ≤10 µm.
 For applications with large temperature changes:
Mount the P-725 onto surfaces that have the same or similar thermal expansion properties
as the P-725.
INFORMATION
 Note that only the thread adapter on the objective side is required when mounting the
P-725 using the mounting holes in the base body.
The following figures serve as examples and can differ from your model.
Fig. 8:
P-725.xxx (example view from above):
Mounting holes (black arrows) and locating holes
(hatched arrows) in the base body
Requirements
 You have read and understood the safety precautions (p. 6).
 The P-725 is not connected to the electronics.
 The flatness of the mounting surface is ≤10 µm.
Tools and accessories
Four M2.5 screws of suitable length (p. 34) for screwing the P-725 to the microscope
Optional: Two suitable locating pins (p. 34) for aligning the P-725
Thread adapter, objective side (refer to "Optional Accessories", p. 11).
Suitable screwdriver
Physik Instrumente (PI ) GmbH & Co. KG, Auf der Roemerstrasse 1, 76228 Karlsruhe, Germany
Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,
Fig. 9:
P-725.1CDE2 (example view from above):
Mounting holes (black arrows) and locating holes
(hatched arrows) in the base body
www.pi.ws
Page 24 / 40

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents