Set
PULSE
DURATION
selector to .1 ms, PERIOD
selector
to
.2 ms,
and
+OUTPUT
AMPLITUDE
control fully cw . (Check
that
the
VARIABLE
controls are at X1 .)
NOTE
Perform steps 5 & 6 only if
output amplifier transistor
0140 and 0180 have bee
replaced
or if it is sus
petted that output pul
aberrations are excessive.
ADJUSTMENT PROCEDURE
Internal Amplitude Moni-
tor Switch (see text)
NOTE
Read
information
under
Adjustment Procedure on
page
3-1
before
pro-
ceeding .
Adjust R1i4 (+ Amplitude
Set) for a + pulse ampli-
tude of 5.2 V .
Monitor the + pulse output
with the Real-Time'Oscillo-
Monitor
the
-
output
scope. Adjust R23 (Period
pulse with the Real-Time
Set) for a + pulse period of
Oscilloscope . Adjust R154
0.2 ms and R90 (Duration
(- Amplitude Set" for a -
Setj for a + pulse duration
pulse amplitude of 5.2 V.
of 0.1 ms .
`~"
o`
`~
~
CU
U
n
R 154
C 143
o
o
,
0
~-
k"
, ._C183
Monitor the + output pulse
through the 10X Attenu-
ator
with
the
Sampling
Oscilloscope . Adjust the +
OUTPUT
AMPLITUDE
control for a 5 V output
pulse
amplitude .
Adjust
C143 1+ Output
Comp
Set
!
for
minimum
aberrations
on
the
dis-
played
pulse at a sweep
rate of 10 ns/div .
NOTE ;
COMPONENTS SHOWN WITH DASHED LINES
ARE LOCATED ON BACK SLOE OF' BOARD,
Monitor
the
-
output
pulse
through
the
10X
Attenuator
with
the
Sampling
Oscilloscope .
Adjust
the
- OUTPUT
AMPL[TUDE control for a
5 V output
pulse ampli-
tude . Adjust C183 (- Out-
put
Comp
Set
1
for
minimum
aberrations on
the displayed pulse at a
sweep rate of 10 ns/div .