3.1.8 Connecting Power Sensors

3.1.8.1
User Manual 1178.5566.02 ─ 07
The R&S NRX supports a wide range of R&S power sensors. See the data sheet for
detailed information.
Depending on the power sensor, you have different choices for connecting power sen-
sors.

Sensor Connectors A to D

See
Chapter 3.2.1.1, "Sensor Connector A and
"Sensor Connectors C and
Suitable for:
USB and LAN power sensors
R&S NRQ6
R&S NRP‑Zxx power sensors
LAN PoE+
NRQ6
Frequency Selective Power Sensor
Host
Interface
LO I/0
Sample Clock I/O (Default: 120 MHz)
Figure 3-1: Setup with an R&S power sensor (example)
1 = Signal source
2 = R&S power sensor
3 = Host Interface connector
4 = R&S NRP‑ZK8
5 = Sensor connector of the R&S NRX
6 = R&S NRX
Use an R&S NRP‑ZK8 cable to connect an R&S power sensor to the R&S NRX. If you
use an R&S NRP‑ZK6 cable, the reference clock and trigger are not supported.
1. 8-pin female connector of R&S NRP‑ZK8:
a) Insert the screw-lock cable connector into the host interface of the R&S power
sensor.
b) Tighten the union nut manually.
2. 8-pin male connector of R&S NRP‑ZK8:
a) Insert this connector into one of the sensor ports of the R&S NRX.
3. Connect the RF connector of the R&S power sensor to the signal source. For
details, see the user manual of the R&S power sensor.
Note: Incorrectly connecting/disconnecting an R&S power sensor can damage the