Computer
Installation
Instructions
Instruction Manual
B. Computer systems includes:
1. Desktop computer
2. Color monitor
3. Printer(s)
4. Interface module
5. Bar code scanner
6. Interconnecting cables
C. Place the computer, monitor, and printer in desired location on a sturdy bench or counter top away from the
RENATRON
II 100 Series Station's wet area (or protected from fluid spray). Place the "Interface Module" near
®
the RENATRON II 100 Series Station location (i.e. top or side).
D. Attach the interconnecting cables as shown in Figure 3.
Connect the communications cable from the COM 1 serial port on the computer to the connector marked
"Computer" on the "Interface Module." Reference the computer manufacturer's manual.
Connect the printer cable from any of the USB ports on the computer to the label printer.
Connect the "Bar code scanner" to the "Scanner" port on the "Interface Module."
Connect the USB cable from any of the USB ports on the computer to the report printer.
E. Connect the RENATRON II 100 Series station's communication cable to the connector marked "Comm.
Port" on the back of the RENATRON II 100 Series Station to the lowest "port" number connector available on the
"Interface Module." Reference Figure 3.
F. Refer to the computer manufacturer's manual for instructions for connecting the monitor, keyboard and
mouse.
G. Power on sequence of computer system:
1. Printer(s)
2. Monitor
3. Computer
4. Interface module
5. RENATRON II 100 Series Station(s)
Note – Use a multiple outlet, surge protected, power strip with a single switch if desired, powering everything
on at once.
Note – The computer should be plugged into the "ALWAYS ON" outlet of the power strip, and will have to be
powered On or Off independently of the power strip.
H. Power on the equipment in the order listed in Section "G". The computer will boot into the Windows operating
system, double click on the RENALOG
Note – For instructions on the Automatic Mode of RENATRON II 100 Series operation using RENALOG RM
Dialyzer Reprocessing Data Management System refer to the Instruction Manual P/N: 50090-859.
Note – It is very important to use the proper procedure when turning off the computer. NEVER TURN OFF THE
COMPUTER VIA THE POWER SWITCH. When turning off the computer, exit completely out of RENALOG
RM. At the Windows desktop, click START, then SHUT DOWN, then SHUT DOWN again. Windows will
perform its shut down procedure and turn the computer off. If the power switch on the front of the computer
is used to turn off the computer, corruption to Windows, RENALOG RM, or your data may occur.
RM icon to access the login screen.
®