Cantel Medical MEDIVATORS renatron II 100 Series Instruction Manual page 20

Dialyzer reprocessing system
Table of Contents

Advertisement

Instruction Manual
CAUTION – IT IS THE SOLE RESPONSIBILITY OF THE PRESCRIBING PHYSICIAN TO DETERMINE
THE ACCEPTABLE BLOOD COMPARTMENT REFERENCE VOLUME AND PROGRAM MODE FOR
A PARTICULAR TYPE AND SIZE HOLLOW FIBER DIALYZER. THIS REFERENCE VOLUME AND
PROGRAM MODE MUST BE SET FROM THE FRONT PANEL MANUALLY OR INITIALLY ENTERED
INTO THE DATABASE TO ALLOW THE COMPUTER TO AUTOMATICALLY SET THESE VALUES BEFORE
REPROCESSING A DIALYZER. IT IS STRONGLY RECOMMENDED THAT THE MINIMUM ACCEPTABLE
REFERENCE VOLUME BE AT LEAST 80% OF THE PRE-PROCESS VOLUME.
CAUTION – IT IS IMPORTANT FOR THE OPERATOR TO VERIFY THAT THE BLOOD COMPARTMENT
REFERENCE VOLUME SET ON THE MACHINE MANUALLY OR AUTOMATICALLY IS CORRECT FOR
EACH DIALYZER TO BE REPROCESSED.
E. Reprocessing the Dialyzer on the RENATRON
1. Push the switch labeled "ON." "SEL" will be displayed in the "VOLUME" display window and "– –"
will appear in the "PROGRAM STEP" display window.
2. Push the "RESET" switch and the "MUTE" switch at same time to display the proper program mode in
the "PROGRAM STEP" display window for the dialyzer to be reprocessed, i.e. "00," "CH," "HF."
Note – Pre-clean (PC) mode is disabled when using the RENATRON II 100 Series station in manual operation.
3. Touch and hold the "HOLD TO SET" switch.
4. Adjust the "SET" knob until the "VOLUME" display indicates the physician's prescribed blood
compartment reference volume for the dialyzer to be reprocessed.
5. Touch the "START PROCESS" switch. The RENATRON II 100 Series station will now reprocess the
dialyzer in the manner described in the section General Systems Operation, of this manual.
6. When the "PROCESS COMPLETE" message appears the reprocessed dialyzer is ready to be removed from
the RENATRON II 100 Series station.
7. If a "VOLUME FAIL" or "PRESSURE FAIL" message appears press the "START PROCESS" switch
to repeat the test which was failed. The dialyzer may pass the test when retested by the RENATRON II 100
Series station.
F. Reprocessing the dialyzer on the RENATRON II 100 Series station in automatic operation
1. Push switch labeled "ON." "SEL" will be displayed in the "VOLUME" display window and "--" will appear
in the "PROGRAM STEP" display window.
2. In the automatic reprocessing screen of the RENALOG
on interface module, scan the bar code label on the dialyzer to be reprocessed, scan the bar code label
on the RENATRON II 100 Series station which corresponds to the cycle that is to be run and scan your
authentication label. The order of the labels being scanned does not make a difference.
Note – The two cycles available are the complete reprocessing cycle and the pre-clean cycle. To select a complete
reprocessing cycle, scan the barcode label on the RENATRON II 100 Series station, the dialyzer's bar coded label.
After the third label, your authentication label is scanned, the computer will automatically set the prescribed
blood compartment reference volume and correct program mode for the dialyzer to be reprocessed. To select the
pre-clean option, scan the bar code labeled pre-clean, the dialyzer's bar code label and your authentication label.
Note – Maximum effectiveness of the pre-clean mode is attained in two hours. During this period, if the dialyzer is
not removed from the RENATRON II 100 Series, the machine will vent the dialyzer every 10 seconds to prevent
pressure build up. One label will be printed at the end of the pre-clean cycle warning you that the dialyzer has
been pre-cleaned only and must go through the complete reprocessing cycle before being placed in storage.
CAUTION – THE PRE-CLEAN MODE IS INTENDED TO AUTOMATICALLY STOP THE RENATRON II 100
SERIES STATION AFTER THE RENALIN 100 CLEANING PHASE. THIS ACTION ALLOWS THE OPERATOR
TO REMOVE THE DIALYZER AND SET IT ASIDE FOR AN ADDITIONAL SOAK/DWELL CLEANING PERIOD.
THIS PROCESS HELPS CLEAN DIFFICULT DIALYZERS. PRECLEANED DIALYZERS MUST UNDERGO A
FULL REPROCESSING CYCLE BEFORE USE.
3. Touch the "START PROCESS" switch. The RENATRON II 100 Series station will now reprocess the
dialyzer in the manner described in the section General Systems Operation, of this manual.
4. Push the "RESET" switch on the RENATRON II 100 Series station after a "PROCESS COMPLETE",
"VOLUME FAIL", or "PRESSURE FAIL" message has appeared. The "PROGRAM STEP" display will
indicate step 98. Within a few seconds, the computer will read and reset the RENATRON II 100 Series station
and the last program mode used will be displayed in the "PROGRAM STEP" display. After the computer has
read the data, labels are printed out indicating the dialyzer's status. Place one label on the dialyzer. Additional
labels may be printed if needed.
II 100 Series station in manual operation
®
RM Program, when the green light is illuminated
®